A new capsule for long term monitoring in suspected upper GI-bleeding – first in human use of the HemoPill monitor
Endoscopy · 01.04.2024 · 56 (S 02 S 02) · S164
DOI: 10.1055/s-0044-1783051 ↗Kurz gefasst
Fachartikel in der Zeitschrift Endoscopy, erschienen am 01.04.2024. Verfasst von M. Brand und 1 weiteren Autor:innen.
Abstract (Originalsprache)
Aims
Since 2020, a photometric capsule endoscopy (HemoPill acute) has been available for non-invasive diagnosis of acute bleeding in the upper gastrointestinal tract. For long-term monitoring of a potential source of bleeding, a new version with a fixation option is available. We report on the first application of the new HemoPill monitor capsule (28d monitoring) in a patient with fundus variceal haemorrhage and subsequent fundus variceal occlusion with cyano acrylate.
Methods
The capsule system consists of a photometric capsule, which is attached to an over-the-scope clip by a thread. The system is inserted using a standard gastroscope with a dedicated cap on the tip of the endoscope. The data transmitted by radio is displayed on an accompanying receiver as a HemoPill index (HI). An HI value≥1.0 is considered as active bleeding in the lumen. In our patient, the capsule was fixed in the distal corpus as part of an esophagogastroduodenoscopy (EGD). After discharge the patient was contacted daily by telephone, had daily blood pressure measurements and weekly Hb checks. After 15 days, the position of the capsule and the fundic varices were checked by EGD. After 28 days, the capsule including over-the-scope clip was removed.
Results
The fixation of the capsule (EGD duration 8 minutes), the measurement itself and the patient's handling of the system were free of complications. There were no clinically relevant bleeding events during the 28-day monitoring period (no melena, Hb increase of 1.4 g/dl). There were 10 relevant events (HI value≥1.0) during the measurement period of 28d. Of these, four events lasted less than 10 minutes, and three events less than 60 minutes. Two events lasted more than 90 minutes but did not lead to an emergency endoscopy due to the lack of clinical bleeding signs. At the control endoscopies (d15 and d28), mucosal haemorrhage due to penetrating cyanoacrylate was postulated as a possible cause of the self-limiting haemorrhage. Capsule removal using the over-the-scope clip Remove-System was without complications (procedure time 12 minutes).
Conclusions
The first application of the HemoPill monitor capsule was technically successful and clinically straightforward. The highly sensitive measuring system detects even the smallest haemorrhages, which are not always clinically relevant. For effective monitoring of potential sources of bleeding, threshold values (duration, etc.) must be established in order to avoid unnecessary endoscopies. This requires more comprehensive data from additional patients. Publication History Article published online: 15 April 2024 © 2024. European Society of Gastrointestinal Endoscopy. All rights reserved. Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany
Zitieren
APA
Brand, M., & Meining, A. (2024). A new capsule for long term monitoring in suspected upper GI-bleeding – first in human use of the HemoPill monitor. Endoscopy, 56(S 02 S 02), S164. https://doi.org/10.1055/s-0044-1783051
Vancouver
Brand M, Meining A. A new capsule for long term monitoring in suspected upper GI-bleeding – first in human use of the HemoPill monitor. Endoscopy. 2024;56(S 02 S 02):S164. doi:10.1055/s-0044-1783051
BibTeX
@article{Brand2024,
author = {Brand, M. and Meining, A.},
title = {A new capsule for long term monitoring in suspected upper GI-bleeding – first in human use of the HemoPill monitor},
journal = {Endoscopy},
year = {2024},
volume = {56},
number = {S 02 S 02},
pages = {S164},
doi = {10.1055/s-0044-1783051}
}