Magnetic Balloon-Assisted Colonoscopy in Patients with Prolonged Cecal Intubation Time: A Single-Arm European Multicenter Clinical Investigation
Endoscopy · 01.04.2024 · 56 (S 02 S 02) · S50
DOI: 10.1055/s-0044-1782794 ↗Kurz gefasst
Fachartikel in der Zeitschrift Endoscopy, erschienen am 01.04.2024. Verfasst von A. Repici und 7 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref).
Abstract (Originalsprache)
Aims
Incomplete colonoscopy has been associated with higher risk of post-colonoscopy interval cancer. Colon loop formation is the main risk factor for incomplete colonoscopies. To address these challenges, a marked magnetic balloon technology add-on device was developed to facilitate colonoscope unlooping and progression. This study aims to assess the safety and efficacy of magnetic balloon-assisted colonoscopy in completing prolonged procedures.
Methods
We conducted an open-label, single-arm, prospective, post-market, multicenter study in Italy, Belgium, and Germany. Outpatients undergoing diagnostic or surveillance colonoscopy were eligible if cecal intubation was not achieved within 10 minutes. Patients with angulated and fixed colon curves were excluded. Study technology consists of a balloon catheter that can be inserted on demand in the colonoscope tool channel, filled with a syringe of ferromagnetic fluid, and anchored with an external permanent magnet. Magnetic balloon anchorage stabilizes the scope tip and facilitates easy straightening. Primary endpoint was an incompletion rate≤10%. Rate of serious adverse events was also collected.
Results
Between January and May 2023, a total of 38 patients who experienced an insertion time≥10 minutes with incompleteness of colonoscopy, were included for the interim analysis. Technical success of the magnetic balloon technology was 100%. The cecum was successfully intubated in all 38 patients, achieving a colonoscopy completion rate of 100%, also corresponding to a 0% incompleteness rate (95% CI: 0% – 7.6%). Polyp detection rate was 45% (95% CI: 26% – 71%).
Conclusions
This clinical investigation provides evidence that magnetic balloon-assisted colonoscopy is both safe and effective in completing prolonged colonoscopies. This on-demand technology has the potential to serve as a useful tool for large-scale solution for facilitating colonoscopy completion in a subset of patients at a higher risk of incomplete procedures or adverse events. Publication History Article published online: 15 April 2024 © 2024. European Society of Gastrointestinal Endoscopy. All rights reserved. Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany
Zitieren
APA
Repici, A., Hassan, C., rizkala, T., Spadaccini, M., Fugazza, A., Pellegatta, G., Neumann, H., & Bisschops, R. (2024). Magnetic Balloon-Assisted Colonoscopy in Patients with Prolonged Cecal Intubation Time: A Single-Arm European Multicenter Clinical Investigation. Endoscopy, 56(S 02 S 02), S50. https://doi.org/10.1055/s-0044-1782794
Vancouver
Repici A, Hassan C, rizkala T, Spadaccini M, Fugazza A, Pellegatta G, et al. Magnetic Balloon-Assisted Colonoscopy in Patients with Prolonged Cecal Intubation Time: A Single-Arm European Multicenter Clinical Investigation. Endoscopy. 2024;56(S 02 S 02):S50. doi:10.1055/s-0044-1782794
BibTeX
@article{Repici2024,
author = {Repici, A. and Hassan, C. and rizkala, T. and Spadaccini, M. and Fugazza, A. and Pellegatta, G. and Neumann, H. and Bisschops, R.},
title = {Magnetic Balloon-Assisted Colonoscopy in Patients with Prolonged Cecal Intubation Time: A Single-Arm European Multicenter Clinical Investigation},
journal = {Endoscopy},
year = {2024},
volume = {56},
number = {S 02 S 02},
pages = {S50},
doi = {10.1055/s-0044-1782794}
}
EUVIMED