Aims
Accurate measurement of polyp size during colonoscopy is crucial for informing clinical decisions such as resection techniques and surveillance scheduling. This study aimed to compare polyp size measurement accuracy when using a virtual scale endoscope (VSE) or snare-based polyp size measurement.
Methods
This randomized controlled trial enrolled 221 patients undergoing screening, surveillance or diagnostic outpatient colonoscopies. Study subjects were randomized to have all detected polyps measured for size either using VSE or a snare of known size to estimate the size of each polyp during the colonoscopy. All polyps were measured for reference size directly after their removal from the colon using a digital caliper and before formalin fixation.
Results
93 polyps were included in the VSE group and 102 in the Snare group. VSE demonstrated significantly higher relative accuracy (80.0% [95% CI: 77.0-82.9]) compared to snare-based size estimation (66.4% [95% CI: 62.4-70.5]; p 2mm (13.1% vs. 39.3%) and>3mm (22.6% vs. 55.4%). For diminutive polyps, VSE better prevented misclassification ( 5mm: 21.4% vs. 73.0%; p=). VSE also outperformed snare in measuring within 10% of reference standard size (30.1% vs. 18.6%) and had lower rates of size underestimation (36.5% vs. 65.7%).
Conclusions
Using VSE improves polyp size measurement accuracy during colonoscopy in comparison with snare-based size estimation. In clinical scenarios, VSE reduced misclassifications at clinically relevant size thresholds 2,3 and 5mm which is relevant for adequate choice of polypectomy techniques or when implementing resect and discard strategies. Publication History Article published online: 15 April 2024 © 2024. European Society of Gastrointestinal Endoscopy. All rights reserved. Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany
Zitieren
APA
Djinbachian, R., Taghiakbari, M., Alj, A., Sidani, S., Kiow, J. L. C., Panzini, B., Bouin, M., & Renteln, D. V. (2024). Comparing a virtual scale endoscope and snares for size measurement accuracy of smaller colorectal polyps: A randomized controlled trial. Endoscopy, 56(S 02 S 02), S49-S50. https://doi.org/10.1055/s-0044-1782792
Vancouver
Djinbachian R, Taghiakbari M, Alj A, Sidani S, Kiow JLC, Panzini B, et al. Comparing a virtual scale endoscope and snares for size measurement accuracy of smaller colorectal polyps: A randomized controlled trial. Endoscopy. 2024;56(S 02 S 02):S49-S50. doi:10.1055/s-0044-1782792
BibTeX
@article{Djinbachian2024,
author = {Djinbachian, R. and Taghiakbari, M. and Alj, A. and Sidani, S. and Kiow, J. Liu Chen and Panzini, B. and Bouin, M. and Renteln, D. Von},
title = {Comparing a virtual scale endoscope and snares for size measurement accuracy of smaller colorectal polyps: A randomized controlled trial},
journal = {Endoscopy},
year = {2024},
volume = {56},
number = {S 02 S 02},
pages = {S49--S50},
doi = {10.1055/s-0044-1782792}
}
EUVIMED