Three-Dimensional Analysis of Polyaxial Volar Locking Plate Position for Distal Radius Fracture
Journal of Wrist Surgery · 01.12.2024 · 13 (06) · 484-491
DOI: 10.1055/s-0044-1782238 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Wrist Surgery, erschienen am 01.12.2024. Verfasst von Yusuke Eda und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, Tsukuba Medical Center, Tsukuba, Ibaraki, Japan. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
To avoid screw penetration into the joint when using the polyaxial volar locking plate (VLP) for osteosynthesis of distal radius fractures, it is important to note that the optimal screw insertion angles depending on the plate positions.
Purpose
The purpose of this study was 2-fold: first, to evaluate the differences of the most distal plate position where the screw does not penetrate into the joint in the three-dimensional (3D) radius models; second, to evaluate the relationship between the plate position and the transverse diameter of the distal radius. Patients and Methods Thirty plain X-rays and computed tomography (CT) scans of healthy wrists were evaluated. The transverse diameter was measured on plain X-rays. 3D radius models were reconstructed from CT data. A 3D image of polyaxial VLP was used to investigate the most distal plate position at three different screw insertion angles. The linear distance between the volar articular edge and the plate edge was measured and compared among different screw insertion angles. The correlations between the plate positions and the transverse diameter were also evaluated. In addition, the relationship between the most distal screw place and articular surface was confirmed with one case of distal radius fracture.
Results
The optimal positions relative to the neutral were 2.7 mm proximal in the distal swing and 1.9 mm distal in the proximal swing. The linear distance was significantly correlated with the transverse diameter in each group. It was confirmed that the relationship between the most distal screw place and articular surface was applicable in the actual case.
Conclusion
The results showed that the most distal position of the polyaxial VLP differed depending on the screw insertion angle and became more proximal as the transverse diameter increased. These results may be useful as a reference for preoperative planning. Levels of Evidence III.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedic Surgery, Tsukuba Medical Center, Tsukuba, Ibaraki, Japan.
- Department of Orthopaedic Surgery, Tokyo Medical University Ibaraki Medical Center, Inashiki, Ibaraki, Japan.
- Department of Orthopaedic Surgery, Kikkoman General Hospital, Noda, Chiba, Japan.
- Department of Orthopaedic Surgery, Institute of Medicine, University of Tsukuba, Tsukuba, Ibaraki, Japan.
Zitieren
APA
Eda, Y., Kohyama, S., Ikumi, A., Ishii, T., Yamazaki, M., & Yoshii, Y. (2024). Three-Dimensional Analysis of Polyaxial Volar Locking Plate Position for Distal Radius Fracture. Journal of Wrist Surgery, 13(06), 484-491. https://doi.org/10.1055/s-0044-1782238
Vancouver
Eda Y, Kohyama S, Ikumi A, Ishii T, Yamazaki M, Yoshii Y. Three-Dimensional Analysis of Polyaxial Volar Locking Plate Position for Distal Radius Fracture. Journal of Wrist Surgery. 2024;13(06):484-491. doi:10.1055/s-0044-1782238
BibTeX
@article{Eda2024,
author = {Eda, Yusuke and Kohyama, Sho and Ikumi, Akira and Ishii, Tomoo and Yamazaki, Masashi and Yoshii, Yuichi},
title = {Three-Dimensional Analysis of Polyaxial Volar Locking Plate Position for Distal Radius Fracture},
journal = {Journal of Wrist Surgery},
year = {2024},
volume = {13},
number = {06},
pages = {484--491},
doi = {10.1055/s-0044-1782238}
}
EUVIMED