Nocturnal Polysomnography without Technical Supervision in the Diagnosis of Respiratory Sleep Disorders, Diagnostic Performance of Home and Sleep Laboratory Studies
Sleep Science · 01.09.2023 · 16 (03) · e310-e316
DOI: 10.1055/s-0043-1772809 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sleep Science, erschienen am 01.09.2023. Verfasst von María Florencia Angellotti und 7 weiteren Autor:innen. Beteiligte Einrichtung: Hospital de Clínicas, Universidad de Buenos Aires, Sleep Laboratory, Buenos Aires, Ciudad Autonoma de Buenos Aires, Argentina. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Introduction
Unattended Polysomnography (type 2 PSG) is a procedure for the diagnosis of sleep-disordered breathing (SDB). Published evidence on its performance and efficacy is limited. Available studies reveal a high rate of lost records that could limit its application.
Objective
To assess the efficacy of type 2 PSG and the rate of studies that must be repeated due to critical loss of signals. Methods prospective, descriptive study. Adult patients with suspected SDB were included. Unattended PSG was performed using portable equipment. 75 patients were connected at home and another 75 in the laboratory, without subsequent monitoring. Records were evaluated to determine the percentage of the night with adequate quality for each of the signals, considered as an evaluable signal for = 70% of the total recording time (TRT). The need to repeat the studies was also estimated.
Results
150 patients were recruited; 44% women; age 57.3 ± 15.4 years; BMI 29.4 ± 6.5. EEG and EOG signals were adequate in 149 records. Flow signal by pressure cannula was adequate in 146 and by thermistor in 67.8%. In only one study the signal of both effort bands were inadequate. Oximetry was lost in 4 cases. Ten tracings (6%) met the criteria for repetition; 8 hospital and 2 home.
Conclusions
Acceptable records were obtained in most unattended PSG studies, both at home and in the sleep laboratory. The rate of repetition of studies due to loss of signal was 6%, with failure in SaO2 or in flow signals being the main cause of the indication.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Hospital de Clínicas, Universidad de Buenos Aires, Sleep Laboratory, Buenos Aires, Ciudad Autonoma de Buenos Aires, Argentina.
- Instituto Argentino de Investigación Neurológica, Sleep Laboratory, Buenos Aires, Ciudad Autonoma de Buenos Aires, Argentina.
ORCID-Profile: Ileana Palma
Zitieren
APA
Angellotti, M. F., Grandval, S., Palma, I., Giovini, V., Jaritos, V., Diaz-Lapasini, G., Carnevale, D., & Nogueira, J. F. (2023). Nocturnal Polysomnography without Technical Supervision in the Diagnosis of Respiratory Sleep Disorders, Diagnostic Performance of Home and Sleep Laboratory Studies. Sleep Science, 16(03), e310-e316. https://doi.org/10.1055/s-0043-1772809
Vancouver
Angellotti MF, Grandval S, Palma I, Giovini V, Jaritos V, Diaz-Lapasini G, et al. Nocturnal Polysomnography without Technical Supervision in the Diagnosis of Respiratory Sleep Disorders, Diagnostic Performance of Home and Sleep Laboratory Studies. Sleep Science. 2023;16(03):e310-e316. doi:10.1055/s-0043-1772809
BibTeX
@article{Angellotti2023,
author = {Angellotti, María Florencia and Grandval, Sofía and Palma, Ileana and Giovini, Vanina and Jaritos, Verónica and Diaz-Lapasini, Gastón and Carnevale, Denise and Nogueira, Juan Facundo},
title = {Nocturnal Polysomnography without Technical Supervision in the Diagnosis of Respiratory Sleep Disorders, Diagnostic Performance of Home and Sleep Laboratory Studies},
journal = {Sleep Science},
year = {2023},
volume = {16},
number = {03},
pages = {e310--e316},
doi = {10.1055/s-0043-1772809}
}
EUVIMED