Development of acoustic rhinometry (ARM) – comparison between two measurement probes (prototype (proto-MS) versus conventional probe (conv-MS)) on 6 nasal models.
Laryngo-Rhino-Otologie · 01.05.2023 · 102 (S 02 S 02) · S329
DOI: 10.1055/s-0043-1767567 ↗Kurz gefasst
Fachartikel in der Zeitschrift Laryngo-Rhino-Otologie, erschienen am 01.05.2023. Verfasst von Fiona Hagensieker und 4 weiteren Autor:innen.
Abstract (Originalsprache)
Introduction
ARM is an established procedure for the objective assessment of nasal airway obstruction. Combined with rhinomanometry and rhinoresistometry, it is considered as the gold standard of pre- and postoperative rhinologic functional diagnostics. Problems exist in the accuracy of measurement in the presence of multiple stenoses. The aim of this study is to compare an advanced measurement probe (proto-MS) with the established Rhino ACOUSTIC System (konv-MS) of Happersberger, Germany.
Methods
20 measurements each were performed on 6 nasal models with the konv-MS and the proto-MS. The following parameters were collected: MCA0, MCA1, MCA2 (area [cm2] and distance [cm], volume 1 (0-5cm [cm3]), volume 2 (2-5cm [cm3]). These measured values were compared 1.) between the two measurement probes and 2.) with the reference values of the nose models. The analysis was performed using Excel 2016 and SPSS Statistics 28.
Results
Compared to the reference – the configuration data given to create nose models using 3D printing – there was significantly improved measurement accuracy when using proto-MS versus conv-MS (e.g., MCA2 area in [cm2]: p 0.8).
Discussion
In nasal models with typical clinical nasal deformities with multiple stenoses, clinically relevant improved measurement accuracy was shown with improved reliability for small nasal dimensions. Conv-MS and proto-MS showed overestimated MCA1 and MCA2 area values compared to the reference. Handling and speed of measurement showed optimized results. AiF Arbeitsgemeinschaft industrieller Forschungsvereinigungen ,,Otto von Guericke“ e.V. Publication History Article published online: 12 May 2023 Georg Thieme Verlag Rüdigerstraße 14, 70469 Stuttgart, Germany
Zitieren
APA
Hagensieker, F., Soete, M., Nierkamp, I., Rudack, C., & Beule, A. (2023). Development of acoustic rhinometry (ARM) – comparison between two measurement probes (prototype (proto-MS) versus conventional probe (conv-MS)) on 6 nasal models. Laryngo-Rhino-Otologie, 102(S 02 S 02), S329. https://doi.org/10.1055/s-0043-1767567
Vancouver
Hagensieker F, Soete M, Nierkamp I, Rudack C, Beule A. Development of acoustic rhinometry (ARM) – comparison between two measurement probes (prototype (proto-MS) versus conventional probe (conv-MS)) on 6 nasal models. Laryngo-Rhino-Otologie. 2023;102(S 02 S 02):S329. doi:10.1055/s-0043-1767567
BibTeX
@article{Hagensieker2023,
author = {Hagensieker, Fiona and Soete, Mia and Nierkamp, Inga and Rudack, Claudia and Beule, Achim-Georg},
title = {Development of acoustic rhinometry (ARM) – comparison between two measurement probes (prototype (proto-MS) versus conventional probe (conv-MS)) on 6 nasal models},
journal = {Laryngo-Rhino-Otologie},
year = {2023},
volume = {102},
number = {S 02 S 02},
pages = {S329},
doi = {10.1055/s-0043-1767567}
}
EUVIMED