AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

This trial aimed to validate a CADe system for colorectal polyp detection versus trained endoscopists in an endoscopy blinded set-up with real-time unblinding.

Methods

The CAD-ARTIPOD trial is a prospective multicenter trial conducted in 9 European centers. CADe specifics were published earlier [ 1 ]. Sample size analysis for superiority required 2048 polyps assuming a difference in sensitivity of 2% with endoscopic diagnosis as gold standard. All endoscopists (ADR 20-50%) were blinded for the CADe output. A second observer classified in real time the CADe output in (1) true positive (detected/not detected by the endoscopist) (2) false negative or (3) clinically relevant or irrelevant false positive detection. If a polyp was only detected by CADe, the endoscopist was unblinded for confirmation of the finding. Histological assessment of endoscopically confirmed polyps was performed by two independent histopathologists.

Results

A number of 2080 polyps were collected, 1549 were histologically confirmed polyps. For the primary endpoint superiority was not reached between CADe and endoscopists (sensitivity (Sn) of 0.94 vs 0.96) (non-inferiority margin of 5%, p = 0.0216). False positive rate was 1.8/min. With histology as standard, CADe outperformed the endoscopists (Sn 0.96 versus 0.95 (p=0.03)) and showed an extra detection rate of 4.8% (Fig. [ 1 ]). Fig. 1 Conclusions This manufacturer independent CADe system used in an innovative and competitive study design demonstrated non-inferiority in comparison to endoscopists’ polyp detection. However, with histology as gold standard CADe outperformed endoscopists. Our study also shows that “watching over a shoulder” in itself leads to higher PDR and ADR Publication History Article published online: 14 April 2023 © 2023. European Society of Gastrointestinal Endoscopy. All rights reserved. Georg Thieme Verlag KG Rüdigerstraße 14, 70469 Stuttgart, Germany

Zitieren

APA

Sinonquel, P., Eelbode, T., Pech, O., Wulf, D. D., Dewint, P., Neumann, H., Antonelli, G., Tate, D. J., Lemmers, A., Pilonis, N. D., Kaminski, M. F., Demedts, I., Hassan, C., Roelandt, P., Maes, F., & Bisschops, R. (2023). Clinical validation of a computer-aided detection model for colorectal polyp detection (CAD-ARTIPOD) trial using a second observer and real-time unblinding. Endoscopy, 55(S 02 S 02), S10. https://doi.org/10.1055/s-0043-1765012

Vancouver

Sinonquel P, Eelbode T, Pech O, Wulf DD, Dewint P, Neumann H, et al. Clinical validation of a computer-aided detection model for colorectal polyp detection (CAD-ARTIPOD) trial using a second observer and real-time unblinding. Endoscopy. 2023;55(S 02 S 02):S10. doi:10.1055/s-0043-1765012

BibTeX
@article{Sinonquel2023,
  author = {Sinonquel, P. and Eelbode, T. and Pech, O. and Wulf, D. De and Dewint, P. and Neumann, H. and Antonelli, G. and Tate, D. J. and Lemmers, A. and Pilonis, N. D. and Kaminski, M. F. and Demedts, I. and Hassan, C. and Roelandt, P. and Maes, F. and Bisschops, R.},
  title = {Clinical validation of a computer-aided detection model for colorectal polyp detection (CAD-ARTIPOD) trial using a second observer and real-time unblinding},
  journal = {Endoscopy},
  year = {2023},
  volume = {55},
  number = {S 02 S 02},
  pages = {S10},
  doi = {10.1055/s-0043-1765012}
}
Kostenlos starten →