EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Indication-Specific Adverse Events and Temporal Trends of Lumen-Apposing Metal Stents: A National Postmarketing Surveillance Analysis (2020–2025)

Indikationsspezifische unerwünschte Ereignisse und zeitliche Trends von Lumen-Apposing-Metal-Stents: nationale Postmarketing-Überwachungsanalyse (2020–2025)

Ishaan Vohra iD, Muhammad K Hasan, Harishankar Gopakumar iD, Dennis Yang iD, Deepanshu Jain, Natalie Cosgrove, Maham Hayat, Ranya Ahmed, Abdul Mohammed, Zubair Khan, Carlos-Rodriguez Bertran iD, Mustafa A Arain

Endoscopy International Open · 21.10.2026 · 14 (CP)

DOI: 10.1055/a-2979-6046 ↗

Auf einen Blick

1775 AXIOS-Meldungen aus der FDA-MAUDE-Datenbank zeigten einen Anstieg der Meldungen um 232 %. Bei Gastroenterostomie traten die höchsten Anteile schwerer unerwünschter Ereignisse auf.

Fragestellung

Wie häufig wurden bei AXIOS-Stents indikationsspezifische Geräteprobleme und unerwünschte Ereignisse gemeldet?

Methode

Retrospektive Analyse von 1775 AXIOS-Unerwünschtheitsmeldungen aus der FDA-MAUDE-Datenbank (2020–2025) mit Zuordnung zu Indikation, Geräteproblem und Outcome.

Ergebnisse

Die jährlichen Meldungen stiegen von 144 in 2020 auf 478 in 2025. Indikationen: Pankreasdrainage 46,2 %, biliäre Drainage 35,7 %, Gastroenterostomie 9,5 %. Häufige Geräteprobleme: Positionierung/Malpositionierung 59,0 %, Distalflansch-Aktivierungsfehler 19,4 %.

Einschränkungen

Passiv erfasste Meldedaten ohne operator- oder gerätebezogene Detaildaten; keine prospektive Registerdaten.

Schlussfolgerung der Autor:innen

Die Meldungen zu AXIOS-Unerwünschtheiten nahmen von 2020 bis 2025 zu. Gastroenterostomie war mit der höchsten Belastung schwerer unerwünschter Ereignisse verbunden.

Kernaussagen

  • Meldungen stiegen von 144 auf 478; plus 232 %
  • Positionierung/Malpositionierung war das häufigste Geräteproblem
  • Gastroenterostomie hatte die höchsten Anteile an Mortalität, Perforation und Infektion

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Endoscopy International Open, erschienen am 21.10.2026. Verfasst von Ishaan Vohra und 11 weiteren Autor:innen. Beteiligte Einrichtung: GastroenterologyUniversity of Illinois Chicago College of Medicine at PeoriaPeoriaIllinoisUnited States. Die Arbeit nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Aims The AXIOS stent is a lumen-apposing metal stent widely used for endoscopic ultrasound-guided drainage of pancreatic fluid collections, biliary drainage, and gastroenterostomy. We analyzed device problems and patient adverse events reported to the FDA Manufacturer and User Facility Device Experience (MAUDE) database (2020–2025) with indication-specific risk stratification.

Methods

We performed a retrospective analysis of 1775 AXIOS adverse event reports from the FDA MAUDE database, systematically classified by procedural indication, device problem, and patient outcome. Each report was assigned to a single highest-severity category, and within-indication proportions were calculated.

Results

Annual reports increased by 232%, from 144 in 2020 to 478 in 2025. The indications comprised pancreatic drainage (46.2%), biliary drainage (35.7%), and gastroenterostomy (9.5%). Device problems included positioning/malpositioning (59.0%), distal flange activation failure (19.4%), migration (7.3%), and proximal detachment (4.5%). Gastroenterostomy carried the highest within-indication proportions of mortality (11.8%), perforation (6.5%), and infection (4.1%), exceeding biliary (3.9%, 6.3%, 1.4%) and pancreatic drainage (2.3%, 3.7%, 1.5%).

Conclusions

AXIOS adverse event reporting increased by 232% from 2020 to 2025, reflecting expanding utilization and enhanced surveillance. Gastroenterostomy carried the highest within-indication burden of serious adverse events. The marked rise in distal flange activation failures—coinciding with December 2025 manufacturer field safety notice—demonstrates postmarketing surveillance value and underscores the need for indication-specific risk stratification and prospective registries capturing operator- and device-level detail that passive surveillance cannot provide.

Schlagwörter

Stichwörter der Autor:innen

adverse eventsgastroenterostomylumen-apposing metal stentpatient safetypostmarketing surveillance

Autor:innen und Institutionen

  • GastroenterologyUniversity of Illinois Chicago College of Medicine at PeoriaPeoriaIllinoisUnited States.
  • Center for Interventional EndoscopyAdventHealth OrlandoOrlandoFloridaUnited States.
  • GastroenterologyBrigham and Women's HospitalBostonMassachusettsUnited States.
  • MedicineUniversity of Central FloridaOrlandoFloridaUnited States.

ORCID-Profile: Ishaan Vohra, Harishankar Gopakumar, Dennis Yang, Carlos-Rodriguez Bertran

Zitieren

APA

Vohra, I., Hasan, M. K., Gopakumar, H., Yang, D., Jain, D., Cosgrove, N., Hayat, M., Ahmed, R., Mohammed, A., Khan, Z., Bertran, C. R., & Arain, M. A. (2026). Indication-Specific Adverse Events and Temporal Trends of Lumen-Apposing Metal Stents: A National Postmarketing Surveillance Analysis (2020–2025). Endoscopy International Open, 14(CP). https://doi.org/10.1055/a-2979-6046

Vancouver

Vohra I, Hasan MK, Gopakumar H, Yang D, Jain D, Cosgrove N, et al. Indication-Specific Adverse Events and Temporal Trends of Lumen-Apposing Metal Stents: A National Postmarketing Surveillance Analysis (2020–2025). Endoscopy International Open. 2026;14(CP). doi:10.1055/a-2979-6046

BibTeX
@article{Vohra2026,
  author = {Vohra, Ishaan and Hasan, Muhammad K and Gopakumar, Harishankar and Yang, Dennis and Jain, Deepanshu and Cosgrove, Natalie and Hayat, Maham and Ahmed, Ranya and Mohammed, Abdul and Khan, Zubair and Bertran, Carlos-Rodriguez and Arain, Mustafa A},
  title = {Indication-Specific Adverse Events and Temporal Trends of Lumen-Apposing Metal Stents: A National Postmarketing Surveillance Analysis (2020–2025)},
  journal = {Endoscopy International Open},
  year = {2026},
  volume = {14},
  number = {CP},
  doi = {10.1055/a-2979-6046}
}
Kostenlos starten →