EUVIMEDEUVIMED entdecken
Beobachtungsstudie

Safety and feasibility of an ERCP robotic system in patients with common bile duct stones: a propensity score-matched observational study

Dehua Tang iD, Ruyan Yan, Lin Zhou, Yonghua Shen iD, Chun Yang, Peng Yan, Ningjing Gao, Bin Zhang, Shanshan Shen iD, Ye Fu, Yuxuan Cheng, Hao Zhu und 6 weitere

Endoscopy · 01.10.2026 · 58 (10) · 1123-1131

DOI: 10.1055/a-2895-1501 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Endoscopy, erschienen am 01.10.2026. Verfasst von Dehua Tang und 17 weiteren Autor:innen. Beteiligte Einrichtung: Nanjing Drum Tower Hospital, The Affiliated Hospital of Nanjing University Medical School, Nanjing University, Department of Gastroenterology, China,…. Hauptthemen laut MEDLINE: Cholangiopancreatography, Endoscopic Retrograde, Choledocholithiasis, Gallstones, Robotic Surgical Procedures. Die Arbeit nennt 20 Literaturangaben. Gefördert von: Shanghai Aohua Photoelectricity Endoscope Co..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Robotic assistance may mitigate occupational hazards associated with endoscopic retrograde cholangiopancreatography (ERCP). We evaluated the safety, feasibility, and learning curve of a new ERCP robotic system. In this observational study, 26 patients with common bile duct stones underwent robotic-assisted ERCP; 6 underwent biliary cannulation only and 20 underwent stone extraction. For the 20 cases with stone extraction, comparisons were performed using 1:1 propensity score matching (PSM) with conventional ERCP performed by the same operator. The safety outcome was adverse events. Feasibility outcomes included technical success, procedural times, operator radiation exposure, and learning-curve characteristics. After PSM, adverse event rates were similar between groups. Post-ERCP pancreatitis occurred in 15.0% (3/20, robotic) and 10.0% (2/20, conventional) (P >0.99), with no bleeding, perforation, or procedure-related mortality. Biliary cannulation and stone extraction success rates were 100% (20/20) in both groups. Robotic-assisted ERCP had a longer procedure time (36.6 vs. 17.0 minutes; P < 0.001), but similar cannulation and fluoroscopy times. In robotic-assisted procedures, radiation exposure was markedly lower in the control room than in the operation room (0.10 vs. 1.26 μSv; P < 0.001). Learning-curve analysis suggested performance stabilization after 10–12 cases. Robotic-assisted ERCP appeared safe and feasible, with high technical success and substantially reduced operator radiation exposure.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cholangiopancreatography, Endoscopic RetrogradeCholedocholithiasisGallstonesRobotic Surgical ProceduresAgedFemaleHumansMaleMiddle AgedCatheterizationFeasibility StudiesFluoroscopyLearning CurveOccupational ExposureOperative TimePancreatitisPropensity ScoreRadiation Exposure

Autor:innen und Institutionen

  • Nanjing Drum Tower Hospital, The Affiliated Hospital of Nanjing University Medical School, Nanjing University, Department of Gastroenterology, China, Nanjing.
  • Peking University College of Future Technology, Department of Biomedical Engineering, China, Beijing.
  • Shanghai Aohua Photoelectricity Endoscope Co.,Ltd, Department of Research and Development, China, Shanghai.

ORCID-Profile: Dehua Tang, Yonghua Shen, Shanshan Shen, Lei Wang

Förderung

  • Shanghai Aohua Photoelectricity Endoscope Co.

Zitieren

APA

Tang, D., Yan, R., Zhou, L., Shen, Y., Yang, C., Yan, P., Gao, N., Zhang, B., Shen, S., Fu, Y., Cheng, Y., Zhu, H., Ni, M., Zhang, X., Zheng, R., Liu, B., Xie, T., & Wang, L. (2026). Safety and feasibility of an ERCP robotic system in patients with common bile duct stones: a propensity score-matched observational study. Endoscopy, 58(10), 1123-1131. https://doi.org/10.1055/a-2895-1501

Vancouver

Tang D, Yan R, Zhou L, Shen Y, Yang C, Yan P, et al. Safety and feasibility of an ERCP robotic system in patients with common bile duct stones: a propensity score-matched observational study. Endoscopy. 2026;58(10):1123-1131. doi:10.1055/a-2895-1501

BibTeX
@article{Tang2026,
  author = {Tang, Dehua and Yan, Ruyan and Zhou, Lin and Shen, Yonghua and Yang, Chun and Yan, Peng and Gao, Ningjing and Zhang, Bin and Shen, Shanshan and Fu, Ye and Cheng, Yuxuan and Zhu, Hao and Ni, Muhan and Zhang, Xiang and Zheng, Ruhua and Liu, Bingyi and Xie, Tianyu and Wang, Lei},
  title = {Safety and feasibility of an ERCP robotic system in patients with common bile duct stones: a propensity score-matched observational study},
  journal = {Endoscopy},
  year = {2026},
  volume = {58},
  number = {10},
  pages = {1123--1131},
  doi = {10.1055/a-2895-1501}
}
Kostenlos starten →