Pitfall of endoscopic ultrasound-guided hepaticoduodenostomy: Impact of hepatic parenchymal volume during device insertion

Takeshi Ogura, Jun Matsuno iD, Takafumi Kanadani, Junichi Nakamura, Hiroki Nishikawa

Endoscopy · 01.12.2025 · 57 (S 01 S 01) · E1128-E1129

DOI: 10.1055/a-2707-3324 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Endoscopy, erschienen am 01.12.2025. Verfasst von Takeshi Ogura und 4 weiteren Autor:innen. Beteiligte Einrichtung: Endoscopy Center, Osaka Medical and Pharmaceutical University Hospital, Osaka, Japan. Die Arbeit nennt 2 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Endoscopic ultrasound-guided hepaticoduodenostomy (EUS-HDS) can be indicated for isolated right hepatic duct obstruction after unsuccessful endoscopic retrograde cholangiopancreatography (ERCP) [ 1 ] [ 2 ]. Although this technique has clinical impact on selected patients, the technical aspects are not yet established because there are few indications for EUS-HDS itself. Compared with EUS-guided hepaticogastrostomy (HGS), pushing force may be transmitted less effectively to the bile duct, and pushback can occur during device insertion. During EUS-HGS, increasing the angle of EUS can increase the pushing force. During EUS-HDS, however, increasing the angle reduces pushback less effectively. To increase the pushing force, hepatic parenchyma volume might be important as a greater volume may reinforce the pushing force. We herein describe a pitfall of EUS-HDS during stent delivery system insertion. After inserting the echoendoscope into the duodenum, the right hepatic bile duct was identified. Superficial intrahepatic bile duct puncture was performed using a 19G needle, and contrast medium was injected ([ Fig. 1 ]). After successful guidewire deployment, ERCP catheter insertion was attempted but was unsuccessful because the pushing force deflected to the right side ([ Fig. 2 ]). Therefore, tract dilation was performed using a drill dilator ([ Fig. 3 ]). Attempted insertion of a stent delivery system (5.9 Fr, HANARO Benefit; Boston Scientific, Marlborough, Massachusetts) into the biliary tract failed for the same reason ([ Fig. 4 ]). We then performed the double-guidewire technique to fit the axis and improve pushing ability, which enabled insertion of the stent delivery system and subsequent stent deployment ([ Fig. 5 ], [ Video 1 ]). As free air was observed after the procedure, the patient was placed under conservative treatment for four days. Fig. 1 Cholangiography confirms the superficial right hepatic duct puncture. Fig. 2 During endoscopic retrograde cholangiopancreatography catheter insertion, the pushing force deflects to the right side (arrow). Fig. 3 Tract dilation is performed using the drill dilator. Fig. 4 Insertion of the stent delivery system failed due to the pushing force deflecting to the right side (arrow). Fig. 5 Successful stent deployment after employing the double-guidewire technique. Download Insertion of the stent delivery system failed due to the pushing force deflecting to the right side.Video 1 In conclusion, choosing a puncture route across a large enough hepatic parenchyma volume may enable successful device insertion during EUS-HDS. Endoscopy_UCTN_Code_TTT_1AS_2AD Endoscopy E-Videos https://eref.thieme.de/e-videos E-Videos is an open access online section of the journal Endoscopy , reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high-quality video and are published with a Creative Commons CC-BY license. Endoscopy E-Videos qualify for HINARI discounts and waivers and eligibility is automatically checked during the submission process. We grant 100% waivers to articles whose corresponding authors are based in Group A countries and 50% waivers to those who are based in Group B countries as classified by Research4Life (see: https://www.research4life.org/access/eligibility/ ). This section has its own submission website at https://mc.manuscriptcentral.com/e-videos . Publication History Article published online: 02 October 2025 © 2025. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution License, permitting unrestricted use, distribution, and reproduction so long as the original work is properly cited. (https://creativecommons.org/licenses/by/4.0/). Georg Thieme Verlag KG Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany

Autor:innen und Institutionen

  • Endoscopy Center, Osaka Medical and Pharmaceutical University Hospital, Osaka, Japan.
  • 2nd Department of Internal Medicine, Osaka Medical and Pharmaceutical University, Osaka, Japan.

ORCID-Profile: Jun Matsuno

Zitieren

APA

Ogura, T., Matsuno, J., Kanadani, T., Nakamura, J., & Nishikawa, H. (2025). Pitfall of endoscopic ultrasound-guided hepaticoduodenostomy: Impact of hepatic parenchymal volume during device insertion. Endoscopy, 57(S 01 S 01), E1128-E1129. https://doi.org/10.1055/a-2707-3324

Vancouver

Ogura T, Matsuno J, Kanadani T, Nakamura J, Nishikawa H. Pitfall of endoscopic ultrasound-guided hepaticoduodenostomy: Impact of hepatic parenchymal volume during device insertion. Endoscopy. 2025;57(S 01 S 01):E1128-E1129. doi:10.1055/a-2707-3324

BibTeX
@article{Ogura2025,
  author = {Ogura, Takeshi and Matsuno, Jun and Kanadani, Takafumi and Nakamura, Junichi and Nishikawa, Hiroki},
  title = {Pitfall of endoscopic ultrasound-guided hepaticoduodenostomy: Impact of hepatic parenchymal volume during device insertion},
  journal = {Endoscopy},
  year = {2025},
  volume = {57},
  number = {S 01 S 01},
  pages = {E1128--E1129},
  doi = {10.1055/a-2707-3324}
}
Kostenlos starten →