Reducing applied force in colonoscopy using a novel soft robotic colonoscope: Head-to-head study

Jabed Foyez Ahmed iD, Korn Borvorntanajanya iD, Jialei Shi, Enrico Franco, Ara Darzi, Ferdinando Rodriguez Baena, Nisha Patel

Endoscopy International Open · 24.07.2025 · 13 (CP)

DOI: 10.1055/a-2641-5827 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Endoscopy International Open, erschienen am 24.07.2025. Verfasst von Jabed Foyez Ahmed und 6 weiteren Autor:innen. Beteiligte Einrichtung: The Hamlyn Centre for Robotic Surgery, Imperial College London, London, United Kingdom of Great Britain and Northern Ireland. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Engineering and Physical Sciences Research Council.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Current colonoscopies have a recognized limitation. Manual pushing and pulling required by operators provides gross movement in the bowel. Reported pain, therefore, is likely due to the manual force applied. Implementing novel steering techniques with a soft growing robotic system can potentially overcome challenges such as fine control, precise steering, and capability to expand treatment options for complex therapies. This study assessed a novel controlled-growing soft robot compared with a standard colonoscope in terms of force exhibited on a model bowel wall and its clinical implications. A head-to-head study using a hybrid colon phantom of the left colon was undertaken. Both the novel soft-robot and standard colonoscope were passed through the phantom by endoscopists. Multiple passes were undertaken in the phantom with both colonoscopy methods with force values recorded at two points (rectum and sigmoid) with pressure sensors. Nine clinical endoscopists (4M:5F, 5 non-expert, 4 expert) were recruited. Average force with the novel robot was 0.25N (rectum) and 0.19N (sigmoid). Average force applied with standard colonoscopy was 2.82N (rectum) and 1.45N (sigmoid). This study demonstrated an improvement in force with the novel soft robot compared with a standard colonoscope. This suggests the possibility of more comfortable colonoscopy for patients. Currently time taken is longer with the novel robot, which is attributable to the learning curve and improves in subsequent passes. Further work will be undertaken in a complete colon model with aspirations to reach in-vivo experiments.

Schlagwörter

Stichwörter der Autor:innen

Endoscopy Lower GI TractPerformance and complicationsQuality and logistical aspectsQuality managementTraining

Autor:innen und Institutionen

  • The Hamlyn Centre for Robotic Surgery, Imperial College London, London, United Kingdom of Great Britain and Northern Ireland.
  • Gastroenterology, Imperial College Healthcare NHS Trust, London, United Kingdom of Great Britain and Northern Ireland.

ORCID-Profile: Jabed Foyez Ahmed, Korn Borvorntanajanya

Förderung

  • Engineering and Physical Sciences Research Council

Zitieren

APA

Ahmed, J. F., Borvorntanajanya, K., Shi, J., Franco, E., Darzi, A., Baena, F. R., & Patel, N. (2025). Reducing applied force in colonoscopy using a novel soft robotic colonoscope: Head-to-head study. Endoscopy International Open, 13(CP). https://doi.org/10.1055/a-2641-5827

Vancouver

Ahmed JF, Borvorntanajanya K, Shi J, Franco E, Darzi A, Baena FR, et al. Reducing applied force in colonoscopy using a novel soft robotic colonoscope: Head-to-head study. Endoscopy International Open. 2025;13(CP). doi:10.1055/a-2641-5827

BibTeX
@article{Ahmed2025,
  author = {Ahmed, Jabed Foyez and Borvorntanajanya, Korn and Shi, Jialei and Franco, Enrico and Darzi, Ara and Baena, Ferdinando Rodriguez and Patel, Nisha},
  title = {Reducing applied force in colonoscopy using a novel soft robotic colonoscope: Head-to-head study},
  journal = {Endoscopy International Open},
  year = {2025},
  volume = {13},
  number = {CP},
  doi = {10.1055/a-2641-5827}
}
Kostenlos starten →