Thermal changes in guidewires used during endoscopic retrograde cholangiopancreatography during electrical conductance: Ex-vivo safety experiment
Endoscopy International Open · 14.03.2025 · 13 (CP)
DOI: 10.1055/a-2536-7997 ↗Kurz gefasst
Fachartikel in der Zeitschrift Endoscopy International Open, erschienen am 14.03.2025. Verfasst von Rahul Karna und 4 weiteren Autor:innen. Beteiligte Einrichtung: Division of Gastroenterology, Hepatology & Nutrition, University of Minnesota Medical Center, Minneapolis, United States. Die Arbeit nennt 5 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Risks of pancreaticobiliary tissue damage secondary to electrical conduction along cannulation wires in the vicinity of electrocautery have rarely been independently studied and remain mostly a theoretical issue. We aimed to evaluate heat generated by commercially available guidewires in the setting of currents used for sphincterotomy to assess their safety during endoscopic procedures. We tested heat generation from 11 common guidewires used in endoscopic retrograde cholangiopancreatography using an electrosurgical generator in combination with a thermocouple temperature probe in an ex vivo model. Thermal changes during electrical conductance were nominal except for minimally wrapped 0.018” wires where the maximum changes in temperature were 4.9°C with ENDO CUT Q settings and 6.8°C with ENDO CUT I settings. Also, the Glidewire 0.018” and the Visiglide 0.025” produced visible sparks along their distal ends with defects in insulation found later under microscopic evaluation. In our ex-vivo study, minimal heat was generated via electrical conductivity despite direct current, suggesting negligible risk of thermal injury during sphincterotomy.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Gastroenterology, Hepatology & Nutrition, University of Minnesota Medical Center, Minneapolis, United States.
- Department of Medicine, Division of Gastroenterology, Hepatology & Nutrition, Thomas Jefferson University, Philadelphia, United States.
- Department of Biomedical Engineering, University of Minnesota, Minneapolis, United States.
ORCID-Profile: Rahul Karna
Zitieren
APA
Karna, R., McDonald, N. M., Weiner, J., Netoff, T. I., & Amateau, S. K. (2025). Thermal changes in guidewires used during endoscopic retrograde cholangiopancreatography during electrical conductance: Ex-vivo safety experiment. Endoscopy International Open, 13(CP). https://doi.org/10.1055/a-2536-7997
Vancouver
Karna R, McDonald NM, Weiner J, Netoff TI, Amateau SK. Thermal changes in guidewires used during endoscopic retrograde cholangiopancreatography during electrical conductance: Ex-vivo safety experiment. Endoscopy International Open. 2025;13(CP). doi:10.1055/a-2536-7997
BibTeX
@article{Karna2025,
author = {Karna, Rahul and McDonald, Nicholas M and Weiner, Jacob and Netoff, Théodon I and Amateau, Stuart K},
title = {Thermal changes in guidewires used during endoscopic retrograde cholangiopancreatography during electrical conductance: Ex-vivo safety experiment},
journal = {Endoscopy International Open},
year = {2025},
volume = {13},
number = {CP},
doi = {10.1055/a-2536-7997}
}