Impact of standard enhancement settings of endoscopy systems on performance of endoscopic artificial intelligence systems

Martijn R Jong, Carolus H J Kusters, Querijn N E van Bokhorst, Jelmer B Jukema, Rixta A H van Eijck van Heslinga, Kiki N Fockens, Britt B S L Houwen, Tim J M Jaspers, Tim G W Boers, Manon van der Vlugt, Evelien Dekker iD, Fons van der Sommen und 4 weitere

Endoscopy · 2025-06 · 57 (06) · 602-610

DOI: 10.1055/a-2530-1845 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Endoscopy, online erschienen am 28.02.2025. Verfasst von Martijn R Jong und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Gastroenterology and Hepatology, Amsterdam Gastroenterology, Endocrinology and Metabolism, Amsterdam UMC, University of Amsterdam, Amst…. Hauptthemen laut MEDLINE: Artificial Intelligence, Colonic Polyps, Barrett Esophagus, Diagnosis, Computer-Assisted. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Nederlandse Organisatie voor Wetenschappelijk Onderzoek.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Artificial intelligence (AI) systems in endoscopy are predominantly developed and tested using high-quality imagery from expert centers. However, their performance may be different when applied in clinical practice, partly due to the diversity in post-processing enhancement settings used in endoscopy units. We evaluated the impact of post-processing enhancement settings on AI performance and tested specific data augmentation strategies to mitigate performance loss.

Methods

We used a computer-aided detection (CADe) system for Barrett’s neoplasia (6223 images, 906 patients) and a computer-aided diagnosis (CADx) system for colorectal polyps (3288 images, 969 patients), both trained on datasets acquired with Olympus equipment and with limited variability in enhancement settings. The CAD systems were then tested across a wide range of test sets, which comprised the same images, but displayed with different enhancement settings. Both CAD systems were then retrained using image enhancement-based data augmentation. The performance of the adjusted CAD systems was evaluated on the same test sets.

Results

Both systems displayed substantial performance variability over a range of enhancement settings (CADe: 83 %–92 % sensitivity, 84 %–91 % specificity; CADx: 78 %–85 % sensitivity, 45 %–63 % specificity). After retraining, variability in sensitivity and specificity was reduced to 2 % (P < 0.001) and 1 %, respectively (P = 0.003) for CADe, and 2 % (P = 0.03) and 8 %, respectively (P = 0.19) for CADx.

Conclusion

The performance of endoscopic AI systems can vary substantially depending on post-processing enhancement settings of the endoscopy unit. Specific data augmentation can mitigate this performance loss.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Artificial IntelligenceColonic PolypsBarrett EsophagusDiagnosis, Computer-AssistedImage EnhancementColorectal NeoplasmsColonoscopyHumansSensitivity and Specificity

Autor:innen und Institutionen

  • Department of Gastroenterology and Hepatology, Amsterdam Gastroenterology, Endocrinology and Metabolism, Amsterdam UMC, University of Amsterdam, Amsterdam, the Netherlands.
  • Department of Electrical Engineering, Eindhoven University of Technology, Eindhoven, Netherlands.

ORCID-Profile: Evelien Dekker, Jacques J G H M Bergman

Förderung

  • Nederlandse Organisatie voor Wetenschappelijk Onderzoek

Zitieren

APA

Jong, M. R., Kusters, C. H. J., van Bokhorst, Q. N. E., Jukema, J. B., van Eijck van Heslinga, R. A. H., Fockens, K. N., Houwen, B. B. S. L., Jaspers, T. J. M., Boers, T. G. W., van der Vlugt, M., Dekker, E., van der Sommen, F., de With, P. H. N., de Groof, A. J., Bergman, J. J. G. H. M., & BONS-AI Consortium (2025). Impact of standard enhancement settings of endoscopy systems on performance of endoscopic artificial intelligence systems. Endoscopy, 57(06), 602-610. https://doi.org/10.1055/a-2530-1845

Vancouver

Jong MR, Kusters CHJ, van Bokhorst QNE, Jukema JB, van Eijck van Heslinga RAH, Fockens KN, et al. Impact of standard enhancement settings of endoscopy systems on performance of endoscopic artificial intelligence systems. Endoscopy. 2025;57(06):602-610. doi:10.1055/a-2530-1845

BibTeX
@article{Jong2025,
  author = {Jong, Martijn R and Kusters, Carolus H J and van Bokhorst, Querijn N E and Jukema, Jelmer B and van Eijck van Heslinga, Rixta A H and Fockens, Kiki N and Houwen, Britt B S L and Jaspers, Tim J M and Boers, Tim G W and van der Vlugt, Manon and Dekker, Evelien and van der Sommen, Fons and de With, Peter H N and de Groof, Albert J and Bergman, Jacques J G H M and BONS-AI Consortium, BONS-AI Consortium},
  title = {Impact of standard enhancement settings of endoscopy systems on performance of endoscopic artificial intelligence systems},
  journal = {Endoscopy},
  year = {2025},
  volume = {57},
  number = {06},
  pages = {602--610},
  doi = {10.1055/a-2530-1845}
}
Kostenlos starten →