CT Motion-Analysis of Implant Loosening in Total Wrist Arthroplasty: A Pilot Study
Journal of Wrist Surgery · 01.06.2026 · 15 (03) · 237-242
DOI: 10.1055/a-2528-0045 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Wrist Surgery, erschienen am 01.06.2026. Verfasst von Daniel Reiser und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedics and Hand Surgery, Faculty of Medicine and Health, Örebro University, Örebro, Sweden. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The Region Örebro län.
Abstract (Originalsprache)
Introduction
Total wrist arthroplasty (TWA) is a motion-preserving treatment option for wrist arthritis. High-precision measurement methods for implant migration such as computed tomography motion-analysis (CTMA) can potentially detect poor implant fixation. The aim of this pilot study was to assess CTMA as a complementary method to diagnose aseptic loosening of TWA. Materials and Methods Three patients with a TWA and wrist pain during activity underwent induced displacement CT (CTMA) with alternated provocations as a complement to plain radiographs.
Results
Two of the three patients had displacement of the carpal component on CTMA. The radial component was stable in all cases. The tool was adapted to clinical routine use.
Conclusions
CT motion-analysis could be a valuable adjunct to plain radiographs in assessing component loosening in TWA.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopedics and Hand Surgery, Faculty of Medicine and Health, Örebro University, Örebro, Sweden.
- Department of Orthopedic Surgery, Mayo Clinic, Rochester, Minnesota.
- Sectra AB, Linköping, Sweden.
ORCID-Profile: Marcus Sagerfors
Förderung
- The Region Örebro län
Zitieren
APA
Reiser, D., Kakar, S., Sandberg, O., Wretenberg, P., & Sagerfors, M. (2026). CT Motion-Analysis of Implant Loosening in Total Wrist Arthroplasty: A Pilot Study. Journal of Wrist Surgery, 15(03), 237-242. https://doi.org/10.1055/a-2528-0045
Vancouver
Reiser D, Kakar S, Sandberg O, Wretenberg P, Sagerfors M. CT Motion-Analysis of Implant Loosening in Total Wrist Arthroplasty: A Pilot Study. Journal of Wrist Surgery. 2026;15(03):237-242. doi:10.1055/a-2528-0045
BibTeX
@article{Reiser2026,
author = {Reiser, Daniel and Kakar, Sanj and Sandberg, Olof and Wretenberg, Per and Sagerfors, Marcus},
title = {CT Motion-Analysis of Implant Loosening in Total Wrist Arthroplasty: A Pilot Study},
journal = {Journal of Wrist Surgery},
year = {2026},
volume = {15},
number = {03},
pages = {237--242},
doi = {10.1055/a-2528-0045}
}
EUVIMED