Vergleichsstudie

Dose-Optimized Image Acquisition Parameters for Neonatal Chest Radiography: A Phantom Study Comparing Computed Radiography and Wireless Digital Radiography Needle Detectors

Kevin Rama iD, Michael Esser iD, Jakob Spogis, Friedrich Wanninger, Bernd Hoberg, Jürgen Frank Schäfer iD

RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren · 01.12.2025 · 197 (12) · 1416-1425

DOI: 10.1055/a-2525-9430 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, erschienen am 01.12.2025. Verfasst von Kevin Rama und 5 weiteren Autor:innen. Beteiligte Einrichtung: Division of Pediatric Radiology, Department of Diagnostic and Interventional Radiology, University Hospital Tübingen, Tübingen, Germany. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiography, Thoracic, Radiation Dosage, Radiographic Image Enhancement. Die Arbeit nennt 13 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

To determine dose-optimized image acquisition parameters for good image quality (IQ) in neonatal chest radiography with a computed radiography (CR) CsBr needle detector vs. a wireless digital radiography (DR) CsI detector using different doses and filters. Physical resolution of the two detectors in unprocessed imaging of a contrast-detail phantom was automatically evaluated. Post-processed chest radiographic imaging of a neonatal phantom was used for Visual Grading Analysis (VGA) by three radiology raters. Different kVp, mAs, and filter settings were used. The VGA score (VGAS) and dose area product (DAP) were used to determine image acquisition parameters and dose levels for good image quality. Pixel data from segments corresponding to visual grading characteristics (VGC) was used to calculate signal-to-noise ratio, contrast-to-noise ratio (CNR), and signal profile curves. These results were compared to the raters’ “ground truth” by Spearman’s correlation. The CR detector had the highest resolution in unprocessed imaging, although this was dependent on a tube voltage of 66 kVp (P < 0.001), and more so than the DR detector. The VGAS showed no significant difference between the CR needle and the DR CsI detectors at the same DAP, or when using standard pediatric filtering of 3.5 mm Al + 0.1 mm Cu (P > 0.05). A lung dose level of 0.017 mSv was needed for good IQ (effective dose (E): 0.010 mSv). This was achievable with different acquisition parameters. Out of 24 segments, only the CNR of bone-to-soft-tissue had a good Spearman’s correlation (ρ > 0.50) to raters’ VGAS (P < 0.0001), mostly due to problems with image registration. The CR needle and DR CsI detectors have comparable IQ in neonatal chest radiography. In this study, an E of approximately 0.010 mSv was needed for good IQ.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingRadiography, ThoracicRadiation DosageRadiographic Image EnhancementWireless TechnologyTomography, X-Ray ComputedNeedlesHumansInfant, NewbornEquipment DesignSensitivity and SpecificitySignal-To-Noise Ratio

Autor:innen und Institutionen

  • Division of Pediatric Radiology, Department of Diagnostic and Interventional Radiology, University Hospital Tübingen, Tübingen, Germany.
  • Agfa Radiology Solutions, Agfa-Gevaert HealthCare GmbH, Peißenberg, Germany.

ORCID-Profile: Kevin Rama, Michael Esser, Jürgen Frank Schäfer

Zitieren

APA

Rama, K., Esser, M., Spogis, J., Wanninger, F., Hoberg, B., & Schäfer, J. F. (2025). Dose-Optimized Image Acquisition Parameters for Neonatal Chest Radiography: A Phantom Study Comparing Computed Radiography and Wireless Digital Radiography Needle Detectors. RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 197(12), 1416-1425. https://doi.org/10.1055/a-2525-9430

Vancouver

Rama K, Esser M, Spogis J, Wanninger F, Hoberg B, Schäfer JF. Dose-Optimized Image Acquisition Parameters for Neonatal Chest Radiography: A Phantom Study Comparing Computed Radiography and Wireless Digital Radiography Needle Detectors. RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren. 2025;197(12):1416-1425. doi:10.1055/a-2525-9430

BibTeX
@article{Rama2025,
  author = {Rama, Kevin and Esser, Michael and Spogis, Jakob and Wanninger, Friedrich and Hoberg, Bernd and Schäfer, Jürgen Frank},
  title = {Dose-Optimized Image Acquisition Parameters for Neonatal Chest Radiography: A Phantom Study Comparing Computed Radiography and Wireless Digital Radiography Needle Detectors},
  journal = {RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren},
  year = {2025},
  volume = {197},
  number = {12},
  pages = {1416--1425},
  doi = {10.1055/a-2525-9430}
}
Kostenlos starten →