Fetal Heart Rate and Amniotic Fluid Volume Measurements with a Home Ultrasound Device

Anat Pardo, Shir Nahum Fridland, Or Lee Rak, Emilie Klochendler Frishman, Hadar Zafrir Danieli, Anat Shmueli, Shiri Barbash-Hazan, Arnon Wiznitzer, Asnat Walfisch, Tomer Sela, Leor Wolff, Eran Hadar

American Journal of Perinatology · 2025-07 · 42 (09) · 1192-1199

DOI: 10.1055/a-2469-0887 ↗

Kurz gefasst

Fachartikel in der Zeitschrift American Journal of Perinatology, online erschienen am 31.12.2024. Verfasst von Anat Pardo und 11 weiteren Autor:innen. Beteiligte Einrichtung: Rabin Medical Center, Helen Schneider Hospital for Women, Petach Tikva, Israel. Hauptthemen laut MEDLINE: Heart Rate, Fetal, Ultrasonography, Prenatal, Amniotic Fluid. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Pulsenmore.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Pulsenmore ES is a self-scanning ultrasound (US) system for remote fetal assessment. It comprises a handheld transducer that serves as a smartphone cradle coupled with an application and clinician's web-viewer dashboard. Recently, a novel capability was added to the system allowing offline fetal heart rate (FHR) and maximal vertical pocket (MVP) measurements. The aim of this study was to evaluate these tools for usability and accuracy. A prospective, non-randomized, non-blinded clinical study design was used. Pulsenmore ES scans were obtained by non-professional laypersons in app-guided (AG) mode (user follows video tutorials in the application) or clinician-guided (CG) mode (user is guided by a health care professional in a real-time telemedicine visit). The scans were stored on a cloud for later interpretation by a health care professional. Each self-scan was immediately followed by a standard US scan performed by a clinician. The asynchronous FHR and MVP measurements made on the AG and CG scans through the designated dashboard were analyzed and compared with the real-time, in-clinic (INC) measurements. The cohort included 28 women. Rates of successful utilization of the Pulsenmore tool for measurement of FHR were 84.7 ± 11.24% of scans made in AG mode and 96.3 ± 6.35% of scans made in CG mode. Corresponding values for MVP were 91.7 ± 2.31% and 95.0 ± 1.73%. FHR accuracy (difference from INC values) was 10.8 ± 7.5 beats per minute (bpm; 7.2%) in AG mode and 5.8 ± 5.1 bpm (4%) in CG mode. MVP accuracy was 1.3 ± 1.4 cm (22%) and 0.9 ± 0.8 cm (14%), respectively. Sensitivity (87.5% and 100% in AG and CG modes, respectively) and specificity (95% and 95.5% in AG and CG modes, respectively) were established for MVP. FHR and MVP measurements obtained from scans captured by the self-operated Pulsenmore ES ultrasound platform are highly accurate and reliable for clinical use relative to standard INC measurements.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart Rate, FetalUltrasonography, PrenatalAmniotic FluidHumansFemalePregnancyProspective StudiesAdultTelemedicineSmartphoneYoung AdultMobile Applications

Autor:innen und Institutionen

  • Rabin Medical Center, Helen Schneider Hospital for Women, Petach Tikva, Israel.
  • Faculty of Medical and Health Sciences, Tel Aviv University, Tel Aviv, Israel.
  • Translational Innovation and eHealth Wing, Clalit Health Services, Tel Aviv, Israel.

Förderung

  • Pulsenmore

Zitieren

APA

Pardo, A., Nahum Fridland, S., Rak, O. L., Klochendler Frishman, E., Zafrir Danieli, H., Shmueli, A., Barbash-Hazan, S., Wiznitzer, A., Walfisch, A., Sela, T., Wolff, L., & Hadar, E. (2025). Fetal Heart Rate and Amniotic Fluid Volume Measurements with a Home Ultrasound Device. American Journal of Perinatology, 42(09), 1192-1199. https://doi.org/10.1055/a-2469-0887

Vancouver

Pardo A, Nahum Fridland S, Rak OL, Klochendler Frishman E, Zafrir Danieli H, Shmueli A, et al. Fetal Heart Rate and Amniotic Fluid Volume Measurements with a Home Ultrasound Device. American Journal of Perinatology. 2025;42(09):1192-1199. doi:10.1055/a-2469-0887

BibTeX
@article{Pardo2025,
  author = {Pardo, Anat and Nahum Fridland, Shir and Rak, Or Lee and Klochendler Frishman, Emilie and Zafrir Danieli, Hadar and Shmueli, Anat and Barbash-Hazan, Shiri and Wiznitzer, Arnon and Walfisch, Asnat and Sela, Tomer and Wolff, Leor and Hadar, Eran},
  title = {Fetal Heart Rate and Amniotic Fluid Volume Measurements with a Home Ultrasound Device},
  journal = {American Journal of Perinatology},
  year = {2025},
  volume = {42},
  number = {09},
  pages = {1192--1199},
  doi = {10.1055/a-2469-0887}
}
Kostenlos starten →