Prospective single-arm trial on feasibility and safety of an endoscopic robotic system for colonic endoscopic submucosal dissection
Endoscopy · 01.03.2025 · 57 (03) · 240-246
DOI: 10.1055/a-2411-0892 ↗Kurz gefasst
Klinische Studie in der Zeitschrift Endoscopy, erschienen am 01.03.2025. Verfasst von Philip W Y Chiu und 8 weiteren Autor:innen. Beteiligte Einrichtung: Surgery, The Chinese University of Hong Kong, Hong Kong, Hong Kong. Hauptthemen laut MEDLINE: Endoscopic Mucosal Resection, Robotic Surgical Procedures, Colorectal Neoplasms, Colonoscopy. Die Arbeit wurde bisher 24-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: EndoMaster Pte Co Ltd.
Abstract (Originalsprache)
Background
The development of the EndoMaster “Endoluminal Access Surgical Efficacy” (EASE) system aims to enhance the safety and efficacy of colonic endoscopic submucosal dissection (ESD) through two flexible robotic arms. This is the first clinical trial to evaluate the performance of colorectal ESD using EndoMaster.
Method
Patients with early mucosal colorectal neoplasia that was not suitable for en bloc resection with snare-based techniques were recruited. The EndoMaster EASE robotic system consisted of an independently designed flexible robotic platform with two robotic arms. The primary outcome was the complete resection rate using EndoMaster. Secondary outcomes included operating time, hospital stay, procedure-related complications, and oncologic outcomes.
Results
43 patients underwent robotic ESD, with a median robotic dissection time of 49 minutes. The technical success rate was 86.0%, while en bloc resection rate among cases with technical success was 94.6%. The complete resection rate was 83.8% and the median size of specimen was 35 mm (range 15–90 mm). The median hospital stay was 2 (range 1–7) days and there was one delayed bleed after 4 days, which was controlled endoscopically. One patient sustained perforation during the procedure, which was completely closed using clips without sequelae. Two patients were treated by salvage surgery.
Conclusion
This first clinical trial confirmed the safety and efficacy of performing colorectal ESD using the EndoMaster EASE robotic system.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Surgery, The Chinese University of Hong Kong, Hong Kong, Hong Kong.
- Institute of Digestive Disease and State Key Laboratory of GI Diseases, The Chinese University of Hong Kong, Hong Kong, Hong Kong.
- Multi-scale Medical Robotics Center, InnoHK, Hong Kong, Hong Kong.
- School of Mechanical and Aerospace Engineering, Nanyang Technological University, Singapore, Singapore.
- Gastroenterology and Hepatology, National University of Singapore Yong Loo Lin School of Medicine, Singapore, Singapore.
ORCID-Profile: Philip W Y Chiu, Hon Chi Yip, Shannon M Chan, Ho Shing Louis Lau
Förderung
- EndoMaster Pte Co Ltd
Zitieren
APA
Chiu, P. W. Y., Yip, H. C., Chu, S., Chan, S. M., Lau, H. S. L., Tang, R. S. Y., Phee, S. J., Ho, K. Y., & Ng, S. S. M. (2025). Prospective single-arm trial on feasibility and safety of an endoscopic robotic system for colonic endoscopic submucosal dissection. Endoscopy, 57(03), 240-246. https://doi.org/10.1055/a-2411-0892
Vancouver
Chiu PWY, Yip HC, Chu S, Chan SM, Lau HSL, Tang RSY, et al. Prospective single-arm trial on feasibility and safety of an endoscopic robotic system for colonic endoscopic submucosal dissection. Endoscopy. 2025;57(03):240-246. doi:10.1055/a-2411-0892
BibTeX
@article{Chiu2025,
author = {Chiu, Philip W Y and Yip, Hon Chi and Chu, Simon and Chan, Shannon M and Lau, Ho Shing Louis and Tang, Raymond S Y and Phee, Soo Jay and Ho, Khek Yu and Ng, Simon Siu Man},
title = {Prospective single-arm trial on feasibility and safety of an endoscopic robotic system for colonic endoscopic submucosal dissection},
journal = {Endoscopy},
year = {2025},
volume = {57},
number = {03},
pages = {240--246},
doi = {10.1055/a-2411-0892}
}
EUVIMED