Technical failure during colorectal endoscopic full-thickness resection: the “through thick and thin” study
Endoscopy · 01.11.2024 · 56 (11) · 831-839
DOI: 10.1055/a-2328-4753 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Endoscopy, erschienen am 01.11.2024. Verfasst von Giulia Gibiino und 25 weiteren Autor:innen. Beteiligte Einrichtung: Gastroenterology and Digestive Endoscopy Units, Morgagni - Pierantoni Hospital, Forlì, and Maurizio Bufalini Hosptial, Cesena, Italy. Hauptthemen laut MEDLINE: Colorectal Neoplasms, Endoscopic Mucosal Resection. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 29 Literaturangaben.
Abstract (Originalsprache)
Background
Endoscopic full-thickness resection (EFTR) is an effective and safe technique for nonlifting colorectal lesions. Technical issues or failures with the full-thickness resection device (FTRD) system are reported, but there are no detailed data. The aim of our study was to quantify and classify FTRD technical failures.
Methods
We performed a retrospective study involving 17 Italian centers with experience in advanced resection techniques and the required devices. Each center shared and classified all prospectively collected consecutive failures during colorectal EFTR using the FTRD from 2018 to 2022. The primary outcome was the technical failure rate and their classification; secondary outcomes included subsequent management, clinical success, and complications.
Results
Included lesions were mainly recurrent (52 %), with a mean (SD) dimension of 18.4 (7.5) mm. Among 750 EFTRs, failures occurred in 77 patients (35 women; mean [SD] age 69.4 [8.9] years). A classification was proposed: type I, snare noncutting (53 %); type II, clip misdeployment (31 %); and type III, cap misplacement (16 %). Among endoscopic treatments completed, rescue endoscopic mucosal resection was performed in 57 patients (74 %), allowing en bloc and R0 resection in 71 % and 64 %, respectively. The overall adverse event rate was 27.3 %. Pooled estimates for the rates of failure, complications, and rescue endoscopic therapy were similar for low and high volume centers (P = 0.08, P = 0.70, and P = 0.71, respectively).
Conclusions
Colorectal EFTR with the FTRD is a challenging technique with a non-negligible rate of technical failure and complications. Experience in rescue resection techniques and multidisciplinary management are mandatory in this setting.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Gastroenterology and Digestive Endoscopy Units, Morgagni - Pierantoni Hospital, Forlì, and Maurizio Bufalini Hosptial, Cesena, Italy.
- Center for Endoscopic Research Therapeutics and Training (CERTT), Policlinico Agostino Gemelli University, Rome, Italy.
- Digestive Endoscopy Unit, University Campus Bio-Medico, Rome, Italy.
- Digestive Endoscopy Unit, Ospedale Gaetano Fucito, Mercato San Severino, Italy.
- Gastroenterology and Digestive Endoscopy Unit, ARNAS Civico Hospital, Palermo, Italy.
- Gastroenterology and Gastrointestinal Endocopy, Vita-Salute San Raffaele University, Milan, Italy.
ORCID-Profile: Giulia Gibiino, Cecilia Binda, Luigi Giovanni Papparella, Mario Gagliardi, Germana de Nucci, Pietro Fusaroli
Zitieren
APA
Gibiino, G., Binda, C., Papparella, L. G., Spada, C., Andrisani, G., Di Matteo, F. M., Gagliardi, M., Maurano, A., Sferrazza, S., Azzolini, F., Grande, G., de Nucci, G., Cesaro, P., Aragona, G., Cennamo, V., Fusaroli, P., Staiano, T., Soriani, P., Campanale, M., …, & “through thick and thin” study group (2024). Technical failure during colorectal endoscopic full-thickness resection: the “through thick and thin” study. Endoscopy, 56(11), 831-839. https://doi.org/10.1055/a-2328-4753
Vancouver
Gibiino G, Binda C, Papparella LG, Spada C, Andrisani G, Di Matteo FM, et al. Technical failure during colorectal endoscopic full-thickness resection: the “through thick and thin” study. Endoscopy. 2024;56(11):831-839. doi:10.1055/a-2328-4753
BibTeX
@article{Gibiino2024,
author = {Gibiino, Giulia and Binda, Cecilia and Papparella, Luigi Giovanni and Spada, Cristiano and Andrisani, Gianluca and Di Matteo, Francesco Maria and Gagliardi, Mario and Maurano, Attilio and Sferrazza, Sandro and Azzolini, Francesco and Grande, Giuseppe and de Nucci, Germana and Cesaro, Paola and Aragona, Giovanni and Cennamo, Vincenzo and Fusaroli, Pietro and Staiano, Teresa and Soriani, Paola and Campanale, Mariachiara and Di Mitri, Roberto and Pugliese, Francesco and Anderloni, Andrea and Cucchetti, Alessandro and Repici, Alessandro and Fabbri, Carlo and “through thick and thin” study group, “through thick and thin” study group},
title = {Technical failure during colorectal endoscopic full-thickness resection: the “through thick and thin” study},
journal = {Endoscopy},
year = {2024},
volume = {56},
number = {11},
pages = {831--839},
doi = {10.1055/a-2328-4753}
}
EUVIMED