EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Feasibility and safety of a novel plastic stent designed specifically for endoscopic ultrasound-guided pancreatic duct drainage

Soma Fukuda iD, Susumu Hijioka iD, Yoshikuni Nagashio, Yuta Maruki, Akihiro Ohba, Daiki Agarie, Yuya Hagiwara, Hidenobu Hara, Kohei Okamoto, Daiki Yamashige iD, Shin Yagi, Masaru Kuwada und 6 weitere

Endoscopy International Open · 01.06.2024 · 12 (06) · E715-E722

DOI: 10.1055/a-2294-8517 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Endoscopy International Open, erschienen am 01.06.2024. Verfasst von Soma Fukuda und 17 weiteren Autor:innen. Beteiligte Einrichtung: Department of Hepatobiliary and Pancreatic Oncology, National Cancer Center Hospital, Tokyo, Japan. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: National Cancer Center Japan.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and study aims Endoscopic ultrasound-guided pancreatic duct drainage (EUS-PD) is emerging as an effective alternative treatment for obstructive pancreatitis after unsuccessful endoscopic retrograde pancreatography (ERP). However, the high incidence of adverse events associated with EUS-PD (approximately 20%) remains an issue. Recently, we developed a novel plastic stent for EUS-PD, with a radiopaque marker positioned at approximately one-third of the length from the distal end of the stent and side holes positioned exclusively distal to the marker. This study aimed to evaluate the feasibility and safety of using this stent in EUS-PD. Patients and methods We retrospectively reviewed data from 10 patients who underwent EUS-PD with the novel plastic stent at the National Cancer Center Hospital between March 2021 and October 2023. Technical and clinical success, procedure times, adverse events (AEs), recurrent pancreatic duct obstruction (RPO), and time to RPO were assessed.

Results

Of the 10 patients, five had postoperative benign pancreaticojejunal anastomotic strictures and five had malignant pancreatic duct obstruction. The technical and clinical success rates were both 100% (10/10). An AE (self-limited abdominal pain) occurred in one patient (10.0%). Two patients (20.0%) died of their primary disease during the follow-up period (median, 44 days; range, 25–272 days). The incidence of RPO was 10.0% (1/10), and the 3-month non-RPO rate was 83.3%.

Conclusions

The novel plastic stent shows potential as a useful and safe tool in EUS-PD.

Schlagwörter

Stichwörter der Autor:innen

Endoscopic ultrasonographyIntervention EUSPancreasPancreatobiliary (ERCP/PTCD)Strictures

Autor:innen und Institutionen

  • Department of Hepatobiliary and Pancreatic Oncology, National Cancer Center Hospital, Tokyo, Japan.
  • Department of Medicine, Makati Medical Center, Manila, Philippines.
  • Endoscopy Division, National Cancer Center Hospital, Tokyo, Japan.

ORCID-Profile: Soma Fukuda, Susumu Hijioka, Daiki Yamashige

Förderung

  • National Cancer Center Japan

Zitieren

APA

Fukuda, S., Hijioka, S., Nagashio, Y., Maruki, Y., Ohba, A., Agarie, D., Hagiwara, Y., Hara, H., Okamoto, K., Yamashige, D., Yagi, S., Kuwada, M., Chatto, M., Kondo, S., Morizane, C., Ueno, H., Saito, Y., & Okusaka, T. (2024). Feasibility and safety of a novel plastic stent designed specifically for endoscopic ultrasound-guided pancreatic duct drainage. Endoscopy International Open, 12(06), E715-E722. https://doi.org/10.1055/a-2294-8517

Vancouver

Fukuda S, Hijioka S, Nagashio Y, Maruki Y, Ohba A, Agarie D, et al. Feasibility and safety of a novel plastic stent designed specifically for endoscopic ultrasound-guided pancreatic duct drainage. Endoscopy International Open. 2024;12(06):E715-E722. doi:10.1055/a-2294-8517

BibTeX
@article{Fukuda2024,
  author = {Fukuda, Soma and Hijioka, Susumu and Nagashio, Yoshikuni and Maruki, Yuta and Ohba, Akihiro and Agarie, Daiki and Hagiwara, Yuya and Hara, Hidenobu and Okamoto, Kohei and Yamashige, Daiki and Yagi, Shin and Kuwada, Masaru and Chatto, Mark and Kondo, Shunsuke and Morizane, Chigusa and Ueno, Hideki and Saito, Yutaka and Okusaka, Takuji},
  title = {Feasibility and safety of a novel plastic stent designed specifically for endoscopic ultrasound-guided pancreatic duct drainage},
  journal = {Endoscopy International Open},
  year = {2024},
  volume = {12},
  number = {06},
  pages = {E715--E722},
  doi = {10.1055/a-2294-8517}
}
Kostenlos starten →