Low-dose pulsed vs standard pulsed fluoroscopy during ERCP to reduce radiation without change in image quality: Prospective randomized study
Endoscopy International Open · 01.04.2024 · 12 (04) · E554-E560
DOI: 10.1055/a-2284-8656 ↗Kurz gefasst
Fachartikel in der Zeitschrift Endoscopy International Open, erschienen am 01.04.2024. Verfasst von Osman Ali und 9 weiteren Autor:innen. Beteiligte Einrichtung: Gastroenterology, University of Maryland School of Medicine, Baltimore, United States. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIDDK NIH HHS, National Institute of Diabetes and Digestive and Kidney Diseases.
Abstract (Originalsprache)
Background and study aims Endoscopic retrograde cholangiopancreatography (ERCP) poses the risk of radiation exposure (RE) to patients and staff and increases the risk of adverse biological effects such as cataracts, sterility, and cancer. Newer fluoroscopy equipment (C-Arm) provides options to limit radiation in the form of lower radiation dose and frame rate or time-limited “pulsed” settings. However, the impact of lower settings on image quality has not been assessed, and no standard protocol exists for fluoroscopy settings used during ERCP. Patients and methods This was a single-center, double-blind, prospective randomized study of consecutive adult patients undergoing standard-of-care ERCP at a tertiary academic medical center. Patients were randomized into two groups: 1) standard-dose pulsed and 2) low-dose pulsed. Pulsed mode (8 fps) was defined as x-ray exposure either in the manufacturer standard-dose or low-dose settings limited to 3 seconds each time the foot-operated switch was depressed.
Results
Seventy-eight patients undergoing ERCP were enrolled and randomized. No difference in age, gender, or body mass index was found between the two groups. No significant difference in image quality was found between standard-dose and low-dose fluoroscopy P = 0.925). The low-dose group was exposed to significantly less radiation when compared with standard-dose P < 0.05). Fluoroscopy time (minutes) was similar in both groups (2.0 vs 1.9), further suggesting that group assignment had no impact on image quality or procedure time.
Conclusions
Low-dose pulsed fluoroscopy is a reliable method that substantially reduces radiation without compromising image quality or affecting procedure or fluoroscopy times. This underscores the need for standardization in ERCP fluoroscopy settings to limit radiation exposure.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Gastroenterology, University of Maryland School of Medicine, Baltimore, United States.
- Gastroenterology and Hepatology, Carilion Clinic, Roanoke, United States.
- IGS, University of Maryland Institute for Genome Sciences, Silver Spring, United States.
- A&F Environmental Health & Safety, University of Maryland Baltimore, Baltimore, United States.
- Gastroenterology and Hepatology, University of Maryland Baltimore, Baltimore, United States.
- Gastroeneterology, City of Hope Comprehensive Cancer Center, Duarte, United States.
ORCID-Profile: Madeline Alizadeh
Förderung
- NIDDK NIH HHS
- National Institute of Diabetes and Digestive and Kidney Diseases
Zitieren
APA
Ali, O., Kesar, V., Alizadeh, M., Kalachi, K., Twery, B., Wellnitz, N., Kim, R. E., Goldberg, E., Uradomo, L. T., & Darwin, P. E. (2024). Low-dose pulsed vs standard pulsed fluoroscopy during ERCP to reduce radiation without change in image quality: Prospective randomized study. Endoscopy International Open, 12(04), E554-E560. https://doi.org/10.1055/a-2284-8656
Vancouver
Ali O, Kesar V, Alizadeh M, Kalachi K, Twery B, Wellnitz N, et al. Low-dose pulsed vs standard pulsed fluoroscopy during ERCP to reduce radiation without change in image quality: Prospective randomized study. Endoscopy International Open. 2024;12(04):E554-E560. doi:10.1055/a-2284-8656
BibTeX
@article{Ali2024,
author = {Ali, Osman and Kesar, Varun and Alizadeh, Madeline and Kalachi, Kourosh and Twery, Benjamin and Wellnitz, Nicholas and Kim, Raymond Eunho and Goldberg, Eric and Uradomo, Lance T and Darwin, Peter E},
title = {Low-dose pulsed vs standard pulsed fluoroscopy during ERCP to reduce radiation without change in image quality: Prospective randomized study},
journal = {Endoscopy International Open},
year = {2024},
volume = {12},
number = {04},
pages = {E554--E560},
doi = {10.1055/a-2284-8656}
}
EUVIMED