Diagnostic accuracy of artificial intelligence-enabled vectorcardiography versus myocardial perfusion SPECT in patients with suspected or known coronary heart disease
Nuklearmedizin - NuclearMedicine · 01.06.2024 · 63 (03) · 213-218
DOI: 10.1055/a-2263-2322 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Nuklearmedizin - NuclearMedicine, erschienen am 01.06.2024. Verfasst von Simon Aydar und 3 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Radiology, Nuclearmedicine and Molecular Imaging, Heart and Diabetes Center North Rhine-Westphalia, Bad Oeynhausen, Germany. Hauptthemen laut MEDLINE: Myocardial Perfusion Imaging, Sensitivity and Specificity, Tomography, Emission-Computed, Single-Photon, Vectorcardiography. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 6 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Aim
The present study evaluated with myocardial perfusion SPECT (MPS) the diagnostic accuracy of an artificial intelligence-enabled vectorcardiography system (Cardisiography, CSG) for detection of perfusion abnormalities.
Methods
We studied 241 patients, 155 with suspected CAD and 86 with known CAD who were referred for MPS. The CSG was performed after the MPS acquisition. The CSG results (1) p-factor (perfusion, 0: normal, 1: mildly, 2: moderately, 3: highly abnormal) and (2) s-factor (structure, categories as p-factor) were compared with the MPS scores. The CSG system was not trained during the study.
Results
Considering the p-factor alone, a specificity of >78% and a negative predictive value of mostly >90% for all MPS variables were found. The sensitivities ranged from 17 to 56%, the positive predictive values from 4 to 38%. Combining the p- and the s-factor, significantly higher specificity values of about 90% were reached. The s-factor showed a significant correlation (p=0.006) with the MPS ejection fraction.
Conclusions
The CSG system is able to exclude relevant perfusion abnormalities in patients with suspected or known CAD with a specificity and a negative predictive value of about 90% combining the p- and the s-factor. Since it is a learning system there is potential for further improvement before routine use.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Institute of Radiology, Nuclearmedicine and Molecular Imaging, Heart and Diabetes Center North Rhine-Westphalia, Bad Oeynhausen, Germany.
Zitieren
APA
Aydar, S., Knobl, H., Burchert, W., & Lindner, O. (2024). Diagnostic accuracy of artificial intelligence-enabled vectorcardiography versus myocardial perfusion SPECT in patients with suspected or known coronary heart disease. Nuklearmedizin - NuclearMedicine, 63(03), 213-218. https://doi.org/10.1055/a-2263-2322
Vancouver
Aydar S, Knobl H, Burchert W, Lindner O. Diagnostic accuracy of artificial intelligence-enabled vectorcardiography versus myocardial perfusion SPECT in patients with suspected or known coronary heart disease. Nuklearmedizin - NuclearMedicine. 2024;63(03):213-218. doi:10.1055/a-2263-2322
BibTeX
@article{Aydar2024,
author = {Aydar, Simon and Knobl, Hermann and Burchert, Wolfgang and Lindner, Oliver},
title = {Diagnostic accuracy of artificial intelligence-enabled vectorcardiography versus myocardial perfusion SPECT in patients with suspected or known coronary heart disease},
journal = {Nuklearmedizin - NuclearMedicine},
year = {2024},
volume = {63},
number = {03},
pages = {213--218},
doi = {10.1055/a-2263-2322}
}
EUVIMED