Plasma Glucose Concentrations in Different Sampling Tubes Measured on Different Glucose Analysers
Experimental and Clinical Endocrinology & Diabetes · 01.05.2024 · 132 (05) · 260-266
DOI: 10.1055/a-2260-3715 ↗Kurz gefasst
Fachartikel in der Zeitschrift Experimental and Clinical Endocrinology & Diabetes, erschienen am 01.05.2024. Verfasst von Stefan Pleus und 4 weiteren Autor:innen. Beteiligte Einrichtung: Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Lise-Meitner-Straße 8/2, D-89081 Ulm. Hauptthemen laut MEDLINE: Blood Glucose. Die Arbeit nennt 13 Literaturangaben.
Abstract (Originalsprache)
Introduction The German Diabetes Association recommends using sampling tubes with citrate and fluoride additives to diagnose diabetes by oral glucose tolerance test to inhibit glycolysis. The effect of different tubes on measurement results was assessed. Materials and Methods In a first study, an oral glucose tolerance test was performed on 41 participants without anamnestically known diabetes. Venous blood was sampled in two different tubes with citrate/fluoride additives from different manufacturers and one with only lithium-heparin additive. A second study with 42 participants was performed to verify the initial results with an adapted design, in which a third tube with citrate buffer was used, and glucose measurements were performed on two additional devices of another analyser model. Samples were centrifuged either immediately (
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Institut für Diabetes-Technologie, Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm, Lise-Meitner-Straße 8/2, D-89081 Ulm.
Zitieren
APA
Pleus, S., Beil, A., Baumstark, A., Haug, C., & Freckmann, G. (2024). Plasma Glucose Concentrations in Different Sampling Tubes Measured on Different Glucose Analysers. Experimental and Clinical Endocrinology & Diabetes, 132(05), 260-266. https://doi.org/10.1055/a-2260-3715
Vancouver
Pleus S, Beil A, Baumstark A, Haug C, Freckmann G. Plasma Glucose Concentrations in Different Sampling Tubes Measured on Different Glucose Analysers. Experimental and Clinical Endocrinology & Diabetes. 2024;132(05):260-266. doi:10.1055/a-2260-3715
BibTeX
@article{Pleus2024,
author = {Pleus, Stefan and Beil, Alexandra and Baumstark, Annette and Haug, Cornelia and Freckmann, Guido},
title = {Plasma Glucose Concentrations in Different Sampling Tubes Measured on Different Glucose Analysers},
journal = {Experimental and Clinical Endocrinology & Diabetes},
year = {2024},
volume = {132},
number = {05},
pages = {260--266},
doi = {10.1055/a-2260-3715}
}