Bronchoscopy masks mitigate aerosols during upper gastrointestinal endoscopies
Endoscopy International Open · 01.10.2023 · 11 (10) · E1001-E1004
DOI: 10.1055/a-2173-1711 ↗Kurz gefasst
Fachartikel in der Zeitschrift Endoscopy International Open, erschienen am 01.10.2023. Verfasst von Frank Phillips und 4 weiteren Autor:innen. Beteiligte Einrichtung: NIHR Nottingham Biomedical Research Centre, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, United Kingdom. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Medical Research Council, Norgine, Nottingham University Hospitals NHS Trust.
Abstract (Originalsprache)
Background and study aims Upper gastrointestinal endoscopies are considered aerosol-generating procedures (AGP) that risk spread of airborne diseases such as SARS-CoV-2. We aimed to investigate where clinically approved bronchoscopy masks applied to patients during esophagogastroduodenoscopies can mitigate spread of aerosols and droplets. Patients and methods This study included patients undergoing routine upper gastrointestinal endoscopy in a standard endoscopy room and used a particle counter to measure size and number of particles 10 cm from the mouths of 49 patients undergoing upper gastrointestinal endoscopies, of whom 12 wore bronchoscopy masks and 37 did not (controls). Particle counts in the aerosol (≤ 5 µm diameter) and droplet (> 5 µm-diameter) size ranges were measured and averaged over the duration of procedures.
Results
The use of bronchoscopy masks offers a 47% reduction (P = 0.01) in particle count for particles < 5 μm in diameter over the procedure duration (aerosols).
Conclusions
Bronchoscopy masks or similar are a simple, low-cost mitigation technique that can be used during outbreaks of respiratory diseases such as COVID-19 to improve safety and reduce fallow times.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- NIHR Nottingham Biomedical Research Centre, Nottingham University Hospitals NHS Trust and University of Nottingham, Nottingham, United Kingdom.
- Department of Electrical and Electronic Engineering, University of Nottingham, Nottingham, United Kingdom of Great Britain and Northern Ireland.
ORCID-Profile: George S D Gordon
Förderung
- Medical Research Council
- Norgine
- Nottingham University Hospitals NHS Trust
- UK Research and Innovation
Zitieren
APA
Phillips, F., Crowley, J., Warburton, S., Parra-Blanco, A., & Gordon, G. S. D. (2023). Bronchoscopy masks mitigate aerosols during upper gastrointestinal endoscopies. Endoscopy International Open, 11(10), E1001-E1004. https://doi.org/10.1055/a-2173-1711
Vancouver
Phillips F, Crowley J, Warburton S, Parra-Blanco A, Gordon GSD. Bronchoscopy masks mitigate aerosols during upper gastrointestinal endoscopies. Endoscopy International Open. 2023;11(10):E1001-E1004. doi:10.1055/a-2173-1711
BibTeX
@article{Phillips2023,
author = {Phillips, Frank and Crowley, Jane and Warburton, Samantha and Parra-Blanco, Adolfo and Gordon, George S D},
title = {Bronchoscopy masks mitigate aerosols during upper gastrointestinal endoscopies},
journal = {Endoscopy International Open},
year = {2023},
volume = {11},
number = {10},
pages = {E1001--E1004},
doi = {10.1055/a-2173-1711}
}
EUVIMED