Microsurgical Clipping after Failed Contour Device Embolization of an Anterior Communicating Artery Aneurysm: Technical Note
Journal of Neurological Surgery Part A: Central European Neurosurgery · 01.05.2024 · 85 (03) · 319-321
DOI: 10.1055/a-2103-7639 ↗Kurz gefasst
Fallbericht in der Zeitschrift Journal of Neurological Surgery Part A: Central European Neurosurgery, erschienen am 01.05.2024. Verfasst von Christoph J Griessenauer und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurosurgery, Christina Doppler Clinic, Paracelsus Medical University, Salzburg, Austria. Hauptthemen laut MEDLINE: Aneurysm, Ruptured, Embolization, Therapeutic, Endovascular Procedures, Intracranial Aneurysm. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 5 Literaturangaben.
Abstract (Originalsprache)
Background
Endovascular therapy has revolutionized the treatment of cerebral aneurysms in recent years and decades. So-called intrasaccular devices (i.e., Woven EndoBridge [WEB], MicroVention, Aliso Viejo, California, United States; or Contour, Cerus Neurovascular, Fremont, California, United States) are a promising endovascular technology, especially for wide-based aneurysms. However, long-term outcome data are currently particularly scarce for Contour and strategies for failed Contour cases are lacking. Here, we report the feasibility of microsurgical clipping after failed Contour device embolization.
Methods
Feasibility of microsurgical clipping after failed aneurysm embolization with a Contour intrasaccular device was assessed in a patient.
Results
We present the case of a 36-year-old male patient diagnosed with Hunt and Hess grade 1 subarachnoid hemorrhage from an anterior communicating artery aneurysm. The ruptured aneurysm was initially treated with the Contour device. After 3 months, angiographic imaging showed a clear aneurysm residual deemed not endovascularly accessible. The patient was then successfully clipped using microsurgical techniques. The patient was discharged with no neurologic abnormalities.
Conclusion
After thorough bibliographical research, this presents the first published case report of microsurgical clipping after failed embolization with Contour. The main insights gained after clipping were that the Contour does not significantly disturb or hinder clipping. In contrast to coils in aneurysms to be clipped, the Contour can be easily compressed by the clip blades and does not have to be removed. In addition, the Contour had not migrated into the subarachnoid space and there was no abnormal scarring. Clipping appears to be a reasonable treatment strategy for failure of embolization with Contour if endovascular means are not suitable.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Neurosurgery, Christina Doppler Clinic, Paracelsus Medical University, Salzburg, Austria.
- Research Institute of Neurointervention, Paracelsus Medical University, Salzburg, Austria.
ORCID-Profile: Christoph J Griessenauer
Zitieren
APA
Griessenauer, C. J., Killer-Oberpfalzer, M., Beredjiklian, C. M., & Lunzer, M. (2024). Microsurgical Clipping after Failed Contour Device Embolization of an Anterior Communicating Artery Aneurysm: Technical Note. Journal of Neurological Surgery Part A: Central European Neurosurgery, 85(03), 319-321. https://doi.org/10.1055/a-2103-7639
Vancouver
Griessenauer CJ, Killer-Oberpfalzer M, Beredjiklian CM, Lunzer M. Microsurgical Clipping after Failed Contour Device Embolization of an Anterior Communicating Artery Aneurysm: Technical Note. Journal of Neurological Surgery Part A: Central European Neurosurgery. 2024;85(03):319-321. doi:10.1055/a-2103-7639
BibTeX
@article{Griessenauer2024,
author = {Griessenauer, Christoph J and Killer-Oberpfalzer, Monika and Beredjiklian, Carlos M and Lunzer, Manuel},
title = {Microsurgical Clipping after Failed Contour Device Embolization of an Anterior Communicating Artery Aneurysm: Technical Note},
journal = {Journal of Neurological Surgery Part A: Central European Neurosurgery},
year = {2024},
volume = {85},
number = {03},
pages = {319--321},
doi = {10.1055/a-2103-7639}
}
EUVIMED