EUVIMEDEUVIMED entdecken
Evaluationsstudie

The Effect of Cycloplegia on the Biometer for Optical Low-coherence Reflectometry

Konuralp Yakar iD

Klinische Monatsblätter für Augenheilkunde · 01.12.2024 · 241 (12) · 1322-1327

DOI: 10.1055/a-2006-1975 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Klinische Monatsblätter für Augenheilkunde, erschienen am 01.12.2024. Verfasst von Konuralp Yakar. Beteiligte Einrichtung: Ophthalmology, Medicana International Samsun Hospital, Samsun, Turkey. Hauptthemen laut MEDLINE: Biometry. Die Arbeit nennt 35 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The aim of this study was to compare the biometric measurements and different formulas for calculating intraocular lens (IOL) power by using the Lenstar LS900 biometer for optical low-coherence reflectometry before and after induction of cycloplegia in the adult population. Materials and Methods In this cross-sectional study, 168 eyes of 168 healthy volunteers aged 40 – 86 years (59.22 ± 11.57) were included. Biometric measurements, including axial length (AL), anterior chamber depth (ACD), keratometry (K1 and K2), and white-to-white (WTW) were compared using a Lenstar LS900 optical biometer before and after induction of cycloplegia with 1% cyclopentolate. The IOL power was also compared using six different formulas (Barrett Universal II, Haigis, SRK/T, Hoffer Q, Holladay, and SRK-II) for the AcrySof MA60AC IOL before and after induction of cycloplegia.

Results

There were no statistically significant differences in AL, K1 and K2, or WTW measurements before and after induction of cycloplegia. There was a significant increase only in ACD from the biometric parameters after the induction of cycloplegia (p < 0.05). Despite this change, there were no significant changes in IOL power calculations using the six different formulas before and after induction of cycloplegia.

Conclusions

This study demonstrated that IOL power measurements using the Lenstar LS900 can be performed after cycloplegia.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BiometryHumansMiddle AgedAgedFemaleMaleAdultAged, 80 and overReproducibility of ResultsMydriaticsSensitivity and SpecificityLenses, IntraocularEquipment Failure AnalysisCyclopentolateCross-Sectional StudiesEquipment DesignRefraction, OcularInterferometry

Substanzen

MydriaticsCyclopentolate

Autor:innen und Institutionen

  • Ophthalmology, Medicana International Samsun Hospital, Samsun, Turkey.

ORCID-Profile: Konuralp Yakar

Zitieren

APA

Yakar, K. (2024). The Effect of Cycloplegia on the Biometer for Optical Low-coherence Reflectometry. Klinische Monatsblätter für Augenheilkunde, 241(12), 1322-1327. https://doi.org/10.1055/a-2006-1975

Vancouver

Yakar K. The Effect of Cycloplegia on the Biometer for Optical Low-coherence Reflectometry. Klinische Monatsblätter für Augenheilkunde. 2024;241(12):1322-1327. doi:10.1055/a-2006-1975

BibTeX
@article{Yakar2024,
  author = {Yakar, Konuralp},
  title = {The Effect of Cycloplegia on the Biometer for Optical Low-coherence Reflectometry},
  journal = {Klinische Monatsblätter für Augenheilkunde},
  year = {2024},
  volume = {241},
  number = {12},
  pages = {1322--1327},
  doi = {10.1055/a-2006-1975}
}