Agreement Between Ventricular-Arterial Coupling Measured by Carotid Ultrasound and Transesophageal Echocardiography in Cardiac Surgery Patients:A Proof-of-Concept Study

Frederique M de Raat, Esmée C de Boer, Igor W F Paulussen, Joris van Houte, R Arthur Bouwman, Leon J Montenij

Journal of Cardiothoracic and Vascular Anesthesia · 2026

DOI: 10.1053/j.jvca.2025.12.018 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Journal of Cardiothoracic and Vascular Anesthesia, 2026. Verfasst von Frederique M de Raat und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical Engineering, Technical University of Eindhoven, Eindhoven, The Netherlands; Department of Anesthesiology, Pain Therapy and I…. Hauptthemen laut MEDLINE: Echocardiography, Transesophageal, Ultrasonography, Carotid Arteries, Cardiac Surgical Procedures, Carotid Arteries. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Nederlandse Organisatie voor Wetenschappelijk Onderzoek, Technische Universiteit Eindhoven.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVES: Ventricular-arterial coupling (VAC), defined as the ratio of arterial elastance (Ea) to left ventricular end-systolic elastance (Ees), provides valuable insights for optimizing perioperative hemodynamic management. While transthoracic echocardiography and transesophageal echocardiography (TEE) are used to estimate Ees, both modalities require substantial operator expertise and transthoracic echocardiography may be limited by suboptimal imaging conditions, especially in the perioperative setting. Therefore, the authors propose using carotid ultrasound as a noninvasive and more accessible alternative to estimate VAC. Given its close anatomic and physiological relationship with the proximal aorta, the carotid artery reflects central arterial hemodynamics and may serve as a promising surrogate for VAC. This study aimed to evaluate the agreement between carotid ultrasound and TEE in measuring VAC in patients undergoing cardiac surgery. DESIGN: Prospective observational pilot study. SETTING: Operating room. PARTICIPANTS: Twelve adult patients undergoing elective coronary artery bypass grafting surgery. INTERVENTIONS: Ultrasound measurements were obtained both preoperatively and postoperatively using carotid ultrasound and TEE, and the interchangeability between Ea, Ees, and VAC was evaluated with correlation and Bland-Altman analyses. Ea was calculated using the systolic blood pressure and stroke volume, while Ees was estimated using the single-beat method proposed by Chen et al. MEASUREMENTS AND MAIN RESULTS: Strong correlations were observed between carotid ultrasound and TEE measurements for Ea (r = 0.87; 95% confidence interval [CI], 0.73-0.95), Ees (r = 0.82; 95% CI, 0.63-0.92), and VAC (r = 0.79; 95% CI, 0.56-0.90). The estimation of VAC using carotid ultrasound demonstrated clinically acceptable accuracy and precision, with a bias of -0.01 (95% CI, -0.07 to 0.05) and limits of agreement of ±0.30 (range, -0.18 to 0.39). CONCLUSIONS: Carotid ultrasound-derived VAC demonstrated good correlation with VAC measured using TEE in this pilot study. These preliminary results suggest that carotid ultrasound may be useful for investigating VAC at the bedside, but further studies with larger sample sizes are needed to validate these findings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Echocardiography, TransesophagealUltrasonography, Carotid ArteriesCardiac Surgical ProceduresCarotid ArteriesHeart VentriclesVentricular Function, LeftHumansProspective StudiesFemalePilot ProjectsMaleAgedMiddle AgedProof of Concept Study

Stichwörter der Autor:innen

advanced hemodynamic monitoringcarotid ultrasoundleft ventricular elastancesingle-beat assessmentventricular-arterial coupling

Autor:innen und Institutionen

  • Department of Electrical Engineering, Technical University of Eindhoven, Eindhoven, The Netherlands; Department of Anesthesiology, Pain Therapy and Intensive Care, Catharina Hospital, Eindhoven,The Netherlands. Electron…
  • Department of Electrical Engineering, Technical University of Eindhoven, Eindhoven, The Netherlands; Department of Anesthesiology, Pain Therapy and Intensive Care, Catharina Hospital, Eindhoven,The Netherlands.
  • Department of Anesthesiology, Pain Therapy and Intensive Care, Catharina Hospital, Eindhoven,The Netherlands.

Förderung

  • Nederlandse Organisatie voor Wetenschappelijk Onderzoek
  • Technische Universiteit Eindhoven

Zitieren

APA

de Raat, F. M., de Boer, E. C., Paulussen, I. W. F., van Houte, J., Bouwman, R. A., & Montenij, L. J. (2026). Agreement Between Ventricular-Arterial Coupling Measured by Carotid Ultrasound and Transesophageal Echocardiography in Cardiac Surgery Patients:A Proof-of-Concept Study. Journal of Cardiothoracic and Vascular Anesthesia. https://doi.org/10.1053/j.jvca.2025.12.018

Vancouver

de Raat FM, de Boer EC, Paulussen IWF, van Houte J, Bouwman RA, Montenij LJ. Agreement Between Ventricular-Arterial Coupling Measured by Carotid Ultrasound and Transesophageal Echocardiography in Cardiac Surgery Patients:A Proof-of-Concept Study. Journal of Cardiothoracic and Vascular Anesthesia. 2026. doi:10.1053/j.jvca.2025.12.018

BibTeX
@article{deRaat2026,
  author = {de Raat, Frederique M and de Boer, Esmée C and Paulussen, Igor W F and van Houte, Joris and Bouwman, R Arthur and Montenij, Leon J},
  title = {Agreement Between Ventricular-Arterial Coupling Measured by Carotid Ultrasound and Transesophageal Echocardiography in Cardiac Surgery Patients:A Proof-of-Concept Study},
  journal = {Journal of Cardiothoracic and Vascular Anesthesia},
  year = {2026},
  doi = {10.1053/j.jvca.2025.12.018}
}
Kostenlos starten →