Finite Element Analysis of Von Mises Stress in External Fixators for Open Tibial Fractures: A Comparative Study of ASTM F1541‐02 and Tibia‐Based Models in Indonesian Patients
Healthcare Technology Letters · 2026-01 · 13 (1)
DOI: 10.1049/htl2.70068 ↗Kurz gefasst
Fachartikel in der Zeitschrift Healthcare Technology Letters, online erschienen am 11.02.2026. Verfasst von Muhammad Kozin und 7 weiteren Autor:innen. Beteiligte Einrichtung: Research Center for Advanced Materials Badan Riset dan Inovasi Nasional (BRIN) South Tangerang Banten Indonesia. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 185 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Traffic accidents are the leading cause of traumatic fractures worldwide, with tibial fractures being the most common lower‐extremity injury. Open tibial fractures pose significant clinical challenges due to their high risk of infection and non‐union, requiring effective stabilisation. External fixators are widely used for this purpose, but their biomechanical performance must be evaluated under both standardised and patient‐specific conditions. This study presents a finite element analysis of an external fixator tailored to Indonesian patients, using two frameworks: the ASTM F1541‐02 standard protocol and a tibia‐based model derived from patient geometry. Von Mises stress distributions were assessed under axial compression, torsional loading, and four‐point bending. Results showed that stresses predicted by ASTM F1541‐02 were consistently higher than those from the tibia‐based model, particularly under torsion and bending, though all values remained below material yield strength. These findings indicate that the fixator design is safe, while emphasising that tibia‐based modelling provides more physiologically realistic predictions. The study shows the importance of patient‐specific anatomy in computational biomechanics and points to future directions in design optimisation and localized manufacturing.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Research Center for Advanced Materials Badan Riset dan Inovasi Nasional (BRIN) South Tangerang Banten Indonesia.
- Bioengineering and Environmental Sustainability Research Centre University of Liberia Monrovia Montserrado Liberia.
- Department of Biomedical Engineering and Health Sciences Faculty of Electrical Engineering Universiti Teknologi Malaysia Johor Bahru Johor Malaysia.
- Department of Mechanical Engineering Faculty of Engineering Universitas Diponegoro Semarang Central Java Indonesia.
- Department of Mechanical Engineering College of Engineering University of Liberia Monrovia Montserrado Liberia.
- Zenith Allmart Precisindo Sidoarjo East Java Indonesia.
ORCID-Profile: Muhammad Kozin, Muhammad Imam Ammarullah, Abdulfatah Abdu Yusuf, Aghni Ulma Saudi, Siti Amalina Azahra, I Nyoman Jujur, Moch Agus Choiron
Zitieren
APA
Kozin, M., Ammarullah, M. I., Yusuf, A. A., Saudi, A. U., Azahra, S. A., Jujur, I. N., Arrifqi, M. H., & Choiron, M. A. (2026). Finite Element Analysis of Von Mises Stress in External Fixators for Open Tibial Fractures: A Comparative Study of ASTM F1541‐02 and Tibia‐Based Models in Indonesian Patients. Healthcare Technology Letters, 13(1). https://doi.org/10.1049/htl2.70068
Vancouver
Kozin M, Ammarullah MI, Yusuf AA, Saudi AU, Azahra SA, Jujur IN, et al. Finite Element Analysis of Von Mises Stress in External Fixators for Open Tibial Fractures: A Comparative Study of ASTM F1541‐02 and Tibia‐Based Models in Indonesian Patients. Healthcare Technology Letters. 2026;13(1). doi:10.1049/htl2.70068
BibTeX
@article{Kozin2026,
author = {Kozin, Muhammad and Ammarullah, Muhammad Imam and Yusuf, Abdulfatah Abdu and Saudi, Aghni Ulma and Azahra, Siti Amalina and Jujur, I Nyoman and Arrifqi, Muhammad Hirzan and Choiron, Moch Agus},
title = {Finite Element Analysis of Von Mises Stress in External Fixators for Open Tibial Fractures: A Comparative Study of ASTM F1541‐02 and Tibia‐Based Models in Indonesian Patients},
journal = {Healthcare Technology Letters},
year = {2026},
volume = {13},
number = {1},
doi = {10.1049/htl2.70068}
}
EUVIMED