Stability of mass bias correction factors for lead isotope ratio measurements across multiple ICP-MS platforms
Journal of Analytical Atomic Spectrometry · 2026 · 41 (8) · 2921-2934
DOI: 10.1039/d6ja00182c ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Analytical Atomic Spectrometry, 2026. Verfasst von Pranav Seena Prem und 10 weiteren Autor:innen. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 3.0).
Abstract (Originalsprache)
Pb isotope ratios ( R ( 206 Pb/ 208 Pb) = 0.461–9.937 mol mol⁻¹) were measured on eight ICP-MS instruments at two institutes to assess mass-bias correction independence from isotopic composition. ICP-ToF-MS failed to provide unbiased corrections.
Zitieren
APA
Prem, P. S., Rienitz, O., Rosner, M., Oelze, M., Pramann, A., Roethke, A., Schulze, A., Swart, C., Azimzada, A., Schannor, M., & Meermann, B. (2026). Stability of mass bias correction factors for lead isotope ratio measurements across multiple ICP-MS platforms. Journal of Analytical Atomic Spectrometry, 41(8), 2921-2934. https://doi.org/10.1039/d6ja00182c
Vancouver
Prem PS, Rienitz O, Rosner M, Oelze M, Pramann A, Roethke A, et al. Stability of mass bias correction factors for lead isotope ratio measurements across multiple ICP-MS platforms. Journal of Analytical Atomic Spectrometry. 2026;41(8):2921-2934. doi:10.1039/d6ja00182c
BibTeX
@article{Prem2026,
author = {Prem, Pranav Seena and Rienitz, Olaf and Rosner, Martin and Oelze, Marcus and Pramann, Axel and Roethke, Anita and Schulze, Alexander and Swart, Claudia and Azimzada, Agil and Schannor, Mathias and Meermann, Björn},
title = {Stability of mass bias correction factors for lead isotope ratio measurements across multiple ICP-MS platforms},
journal = {Journal of Analytical Atomic Spectrometry},
year = {2026},
volume = {41},
number = {8},
pages = {2921--2934},
doi = {10.1039/d6ja00182c}
}
EUVIMED