Biomechanical effects of length and angle of the subcutaneous tunnel on epidural catheter stability in a porcine model

Jan Pennig, Sarah N Schmidt, Axel Schmutz, Johannes Hell

Scientific Reports · 2026

DOI: 10.1038/s41598-026-71055-7 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Scientific Reports, 2026. Verfasst von Jan Pennig und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology and Critical Care, Medical Center, University of Freiburg, Hugstetter Str. 55, 79106, Freiburg, Germany. Hauptthemen laut MEDLINE: Catheters, Epidural Space, Catheterization. Die Arbeit nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum Freiburg.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Epidural catheter dislocation can lead to severe complications, especially in anticoagulated patients. Fixation techniques vary, and the role of subcutaneous tunneling and the biomechanical effects of length or angle towards dislocation forces of the tunnel remain unknown. This in vitro study evaluated the contribution of length and angle of subcutaneous tunneling (2–5 cm at six various angles relative to dislocation force) compared to different adhesive fixation methods (transparent film dressing, adhesive strips 10–20 cm) on epidural catheter stability. Dislocation forces were measured in a porcine skin model using a motorized traction system and a force transducer. The mean dislodgement force for tunneled catheters was 0.86 ± 0.3 N. This was significantly lower than the forces measured using transparent film dressing (9.89 ± 2.65 N; p < 0.001) or adhesive strips (10 cm: 8.03 ± 2.3 N; 20 cm: 10.92 ± 0.84 N; all p < 0.001). Tunnel length and angle had no significant impact on catheter stability whereas increasing the length of the adhesive strip significantly increased the dislodgement force ( p < 0.001). In conclusion, subcutaneous tunneling of epidural catheters provides limited mechanical benefit on catheter stability under standardized conditions in a porcine model whereas adhesive fixation generates substantially greater resistance to immediate catheter dislocation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CathetersEpidural SpaceCatheterizationAnimalsSwineBiomechanical PhenomenaModels, Animal

Stichwörter der Autor:innen

Catheter stabilizationEpidural catheterEpidural catheter dislocationEpidural catheter fixation

Autor:innen und Institutionen

  • Department of Anesthesiology and Critical Care, Medical Center, University of Freiburg, Hugstetter Str. 55, 79106, Freiburg, Germany.
  • Member of Faculty of Medicine, University of Freiburg, Freiburg, Germany.
  • Department of Anesthesiology and Critical Care, Medical Center, University of Freiburg, Hugstetter Str. 55, 79106, Freiburg, Germany. johannes.hell@uniklinik-freiburg.de.
  • Member of Faculty of Medicine, University of Freiburg, Freiburg, Germany. johannes.hell@uniklinik-freiburg.de.

Förderung

  • Universitätsklinikum Freiburg

Zitieren

APA

Pennig, J., Schmidt, S. N., Schmutz, A., & Hell, J. (2026). Biomechanical effects of length and angle of the subcutaneous tunnel on epidural catheter stability in a porcine model. Scientific Reports. https://doi.org/10.1038/s41598-026-71055-7

Vancouver

Pennig J, Schmidt SN, Schmutz A, Hell J. Biomechanical effects of length and angle of the subcutaneous tunnel on epidural catheter stability in a porcine model. Scientific Reports. 2026. doi:10.1038/s41598-026-71055-7

BibTeX
@article{Pennig2026,
  author = {Pennig, Jan and Schmidt, Sarah N and Schmutz, Axel and Hell, Johannes},
  title = {Biomechanical effects of length and angle of the subcutaneous tunnel on epidural catheter stability in a porcine model},
  journal = {Scientific Reports},
  year = {2026},
  doi = {10.1038/s41598-026-71055-7}
}
Kostenlos starten →