The Scanning CONfoCal Ophthalmoscopy foR DIAbetic eye screening (CONCORDIA) study paper 1
Eye · 2024
DOI: 10.1038/s41433-024-03360-2 ↗Kurz gefasst
Fachartikel in der Zeitschrift Eye, 2024. Verfasst von Peter H Scanlon und 8 weiteren Autor:innen. Beteiligte Einrichtung: Gloucestershire Retinal Research Group (GRRG), Cheltenham General Hospital, Cheltenham, UK. p.scanlon@nhs.net. Hauptthemen laut MEDLINE: Diabetic Retinopathy, Mydriatics, Ophthalmoscopy. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Innovate UK, This study was also supported by NIHR support funds to Gloucestershire Hospitals and unrestricted research grant from Optos, Zeiss and CenterVue SpA..
Abstract (Originalsprache)
Objective
This project was to determine the performance of the Zeiss Clarus 700 (Clarus) and the Optos California (Optos) with staged mydriasis in a Diabetic Eye Screening Programme (DESP).
Methods
Trial participants were recruited from people attending appointments in DESP or Virtual Eye clinics for delayed hospital appointments. Non-mydriatic photographs from the Clarus and Optos cameras were compared to 2-field 45 degrees mydriatic digital photography (the reference standard) and mydriatic photographs compared if the non-mydriatic photos were unassessable (staged mydriasis).
Results
1573 patients were recruited. 76 individuals were withdrawn, leaving 1497 individuals (2993 eyes). For the Clarus and the Optos, the sensitivity for any retinopathy were 94.2% (95% CI: 92.9–95.3%) and 91.9% (95% CI: 90.5–93.2%) with specificities of 87.3% (95% CI: 85.4–89.0%) and 78.1% (95% CI: 75.7–80.3%) respectively. For referable DR the sensitivities for the Clarus and Optos were 86.0% (95% CI: 82.9–88.8%) and 77.6% (95% CI: 73.9–80.9%) with specificities of 92.8% (95% CI: 91.7–93.8%) and 95.4% (95% CI: 94.5–96.2%) respectively. The Clarus and Optos without mydriasis produced 100 (3.3%) and 152 (5.1%) unassessable eyes respectively, and after staged mydriasis 51 (1.7%) and 102 (3.4%) respectively with 52 (1.7%) reference standard images unassessable.
Conclusions
This study reports the performance of the Clarus and the Optos using staged mydriasis in DR screening with wider fields detecting more referable retinopathy peripherally with some reduction in sensitivity centrally for macular lesions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Gloucestershire Retinal Research Group (GRRG), Cheltenham General Hospital, Cheltenham, UK. p.scanlon@nhs.net.
- Nuffield Department of Clinical Neuroscience, University of Oxford, Cheltenham, UK. p.scanlon@nhs.net.
- University of Gloucestershire, Cheltenham, UK. p.scanlon@nhs.net.
- Gloucestershire Retinal Research Group (GRRG), Cheltenham General Hospital, Cheltenham, UK.
- The Royal College of Ophthalmologists' National Ophthalmology Audit, London, UK.
ORCID-Profile: Peter H Scanlon, Marta Gruszka-Goh
Förderung
- Innovate UK
- This study was also supported by NIHR support funds to Gloucestershire Hospitals and unrestricted research grant from Optos, Zeiss and CenterVue SpA.
Zitieren
APA
Scanlon, P. H., Gruszka-Goh, M., Javed, U., Vukic, A., Hapeshi, J., Chave, S., Galsworthy, P., Vallance, S., & Aldington, S. J. (2024). The Scanning CONfoCal Ophthalmoscopy foR DIAbetic eye screening (CONCORDIA) study paper 1. Eye. https://doi.org/10.1038/s41433-024-03360-2
Vancouver
Scanlon PH, Gruszka-Goh M, Javed U, Vukic A, Hapeshi J, Chave S, et al. The Scanning CONfoCal Ophthalmoscopy foR DIAbetic eye screening (CONCORDIA) study paper 1. Eye. 2024. doi:10.1038/s41433-024-03360-2