Comparison of B-Scan ultrasonography, ultra-widefield fundus imaging, and indirect ophthalmoscopy in detecting retinal breaks in cataractous eyes

Ao Miao, Jie Xu, Kai Wei, Peimin Lin, Liangliang Niu, Yumeng Shi, Dongjin Qian, Yi Lu, Yongxiang Jiang, Tianyu Zheng

Eye · 2024

DOI: 10.1038/s41433-024-03093-2 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Eye, 2024. Verfasst von Ao Miao und 9 weiteren Autor:innen. Beteiligte Einrichtung: Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai, China. Hauptthemen laut MEDLINE: Ophthalmoscopy, Cataract, Ultrasonography, Visual Acuity. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: National Natural Science Foundation of China (National Science Foundation of China).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND/OBJECTIVES: To evaluate the diagnostic performance of B-scan kinetic ultrasonography (USG), standard ultra-widefield (UWF) imaging, and indirect ophthalmoscopy (IDO) in retinal break detection in cataractous eyes. SUBJECTS/METHODS: We consecutively enrolled 126 cataract patients (including 246 eyes) with no comorbidities that could decrease best corrected visual acuity (BCVA). Three index tests (USG, nonmydriatic UWF, and mydriatic IDO) were performed preoperatively to screen for retinal breaks. One week after cataract extraction, a dilated IDO examination was repeated for the definitive diagnosis of retinal break as the reference standard. The sensitivity, specificity, Youden index (YI), and predictive values of each index test were calculated according to postoperative ophthalmoscopic findings. A deep-learning nomogram was developed to quantify the risk of retinal break presence using patients' baseline data and findings reported from preoperative ophthalmic tests. RESULTS: Fifty-two eyes (21%) were excluded from appropriate preoperative UWF imaging because of massive lens opacity. The BCVA cutoff point with maximum YI indicating UWF applicability was 0.6 logMAR (YI = 0.3; area under curve [AUC] = 0.7). Among all 246 eyes, preoperative IDO, USG, and UWF showed fair interobserver agreement (all κ > 0.2). According to postoperative IDO findings, the index tests with the highest sensitivity and specificity were USG (100%) and preoperative IDO (99%), respectively. CONCLUSIONS: For cataractous eyes without vision-impairing comorbidities, a BCVA better than 0.6 logMAR (Snellen acuity, 20/80) allows for appropriate nonmydriatic standard UWF imaging. In a high-volume clinic equipped with skilled ophthalmic examiners, screening with USG followed by directed IDO allows the efficient identification of retinal breaks in cataractous eyes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

OphthalmoscopyCataractUltrasonographyVisual AcuityHumansFemaleMaleAgedMiddle AgedProspective StudiesSensitivity and SpecificityPredictive Value of TestsAged, 80 and overFundus OculiCataract ExtractionROC Curve

Autor:innen und Institutionen

  • Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai, China.
  • NHC Key Laboratory of Myopia (Fudan University); Key Laboratory of Myopia, Chinese Academy of Medical Sciences, Shanghai, China.
  • Shanghai Key Laboratory of Visual Impairment and Restoration, Shanghai, China.
  • Bio-Med Big Data Center, CAS Key Laboratory of Computational Biology, Shanghai Institute of Nutrition and Health, University of Chinese Academy of Sciences, Chinese Academy of Sciences, Shanghai, China.
  • Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai, China. yongxiang_jiang@163.com.
  • NHC Key Laboratory of Myopia (Fudan University); Key Laboratory of Myopia, Chinese Academy of Medical Sciences, Shanghai, China. yongxiang_jiang@163.com.

Förderung

  • National Natural Science Foundation of China (National Science Foundation of China)

Zitieren

APA

Miao, A., Xu, J., Wei, K., Lin, P., Niu, L., Shi, Y., Qian, D., Lu, Y., Jiang, Y., & Zheng, T. (2024). Comparison of B-Scan ultrasonography, ultra-widefield fundus imaging, and indirect ophthalmoscopy in detecting retinal breaks in cataractous eyes. Eye. https://doi.org/10.1038/s41433-024-03093-2

Vancouver

Miao A, Xu J, Wei K, Lin P, Niu L, Shi Y, et al. Comparison of B-Scan ultrasonography, ultra-widefield fundus imaging, and indirect ophthalmoscopy in detecting retinal breaks in cataractous eyes. Eye. 2024. doi:10.1038/s41433-024-03093-2

BibTeX
@article{Miao2024,
  author = {Miao, Ao and Xu, Jie and Wei, Kai and Lin, Peimin and Niu, Liangliang and Shi, Yumeng and Qian, Dongjin and Lu, Yi and Jiang, Yongxiang and Zheng, Tianyu},
  title = {Comparison of B-Scan ultrasonography, ultra-widefield fundus imaging, and indirect ophthalmoscopy in detecting retinal breaks in cataractous eyes},
  journal = {Eye},
  year = {2024},
  doi = {10.1038/s41433-024-03093-2}
}
Kostenlos starten →