PIXImus DXA with Different Software Needs Individual Calibration to Accurately Predict Fat Mass
Obesity Research · 2005-09 · 13 (9) · 1558-1565
DOI: 10.1038/oby.2005.191 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Obesity Research, online erschienen am 06.09.2012. Verfasst von Sarah L Johnston und 4 weiteren Autor:innen. Beteiligte Einrichtung: Aberdeen Center for Energy Regulation and Obesity, Obesity and Metabolic Health Division, Rowett Research Institute, Aberdeen, Scotland, UK. sarah.jo…. Hauptthemen laut MEDLINE: Software. Die Arbeit wurde bisher 37-mal zitiert (Crossref) und nennt 15 Literaturangaben.
Abstract (Originalsprache)
Objective
To validate GE PIXImus2 DXA fat mass (FM) estimates by chemical analysis, to compare previously published correction equations with an equation from our machine, and to determine intermachine variation. Research Methods and Procedures : C57BL/6J ( n = 16) and Aston ( n = 14) mice (including ob/ob ), Siberian hamsters ( Phodopus sungorus ) ( n = 15), and bank voles ( Clethrionomys glareolus ) ( n = 37) were DXA scanned postmortem, dried, then fat extracted using a Soxhlet apparatus. We compared extracted FM with DXA‐predicted FM corrected using an equation designed using wild‐type animals from split‐sample validation and multiple regression and two previously published equations. Sixteen animals were scanned on both a GE PIXImus2 DXA in France and a second machine in the United Kingdom.
Results
DXA underestimated FM of obese C57BL/6J by 1.4 ± 0.19 grams but overestimated FM for wild‐type C57BL/6J (2.0 ± 0.11 grams), bank voles (1.1 ± 0.09 grams), and hamsters (1.1 ± 0.13 grams). DXA‐predicted FM corrected using our equation accurately predicted extracted FM (accuracy 0.02 grams), but the other equations did not (accuracy, −1.3 and −1.8 grams; paired Student's t test, p < 0.001). Two similar DXA instruments gave the same FM for obese mutant but not lean wild‐type animals.
Discussion
DXA using the same software could use the same correction equation to accurately predict FM for obese mutant but not lean wild‐type animals. PIXImus machines purchased with new software need validating to accurately predict FM.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Aberdeen Center for Energy Regulation and Obesity, Obesity and Metabolic Health Division, Rowett Research Institute, Aberdeen, Scotland, UK. sarah.johnston@rowett.ac.uk
Zitieren
APA
Johnston, S. L., Peacock, W. L., Bell, L. M., Lonchampt, M., & Speakman, J. R. (2005). PIXImus DXA with Different Software Needs Individual Calibration to Accurately Predict Fat Mass. Obesity Research, 13(9), 1558-1565. https://doi.org/10.1038/oby.2005.191
Vancouver
Johnston SL, Peacock WL, Bell LM, Lonchampt M, Speakman JR. PIXImus DXA with Different Software Needs Individual Calibration to Accurately Predict Fat Mass. Obesity Research. 2005;13(9):1558-1565. doi:10.1038/oby.2005.191
BibTeX
@article{Johnston2005,
author = {Johnston, Sarah L and Peacock, Wendy L and Bell, Lynn M and Lonchampt, Michel and Speakman, John R},
title = {PIXImus DXA with Different Software Needs Individual Calibration to Accurately Predict Fat Mass},
journal = {Obesity Research},
year = {2005},
volume = {13},
number = {9},
pages = {1558--1565},
doi = {10.1038/oby.2005.191}
}
EUVIMED