Fachartikel in der Zeitschrift Chemical & Biomedical Imaging, online erschienen am 06.02.2026. Verfasst von Stephania Rodríguez Muiña und 3 weiteren Autor:innen. Beteiligte Einrichtung: BC Materials, Basque Center for Materials, Applications and Nanostructures. UPV/EHU Parque Científico, BCMaterials, Bizkaia, Leioa E-48940, Spain. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 63 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: HORIZON EUROPE Marie Sklodowska-Curie Actions, NextGenerationEU.
Abstract (Originalsprache)
Electrochemiluminescence (ECL) offers high analytical sensitivity, low optical background, and spatially confined light emission at electrode interfaces, enabling a large global market for bead-based ECL immunoassays. Existing ECL instrumentation remains bulky, costly, and largely confined to centralized laboratories with limited access to true spatially resolved imaging. Here, we present a compact, modular ECL imaging platform that enables low-light, spatially resolved, and temporally synchronized electrochemical–optical measurements. The system integrates a Raspberry Pi 4 controller, a monochrome OV9281 global-shutter camera, and an EmStat4s potentiostat, coordinated through open-source Python software providing sub-100 ms synchronization and full control of camera parameters (resolution, exposure, gain, frame rate, and focus). The complete system, of approximately 21 × 15 × 21 cm, has a component bill of materials of 2337 €, which could be reduced to a few hundred euros by using a custom potentiostat, making it substantially more affordable than commercial ECL instruments. Platform performance was validated using both homogeneous solution-phase and heterogeneous bead-based [Ru(bpy)3]2+/tripropylamine (TPrA) assays on carbon screen-printed electrodes. Synchronized cyclic voltammetry and chronoamperometry captured ECL transient behavior, including onset, peak emission, and decay dynamics. Quantitative bead-based assays showed excellent linearity with low-picomolar range limits of detection, while spatially resolved analysis enabled multiplexed measurements on a single electrode via a region-of-interest readout. These results demonstrate the importance of controlled and optimized camera parameters for quantitative ECL imaging and establish this compact, low-cost, and modular platform as a practical alternative for spatially resolved ECL biosensing and multiplex point-of-care applications.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- BC Materials, Basque Center for Materials, Applications and Nanostructures. UPV/EHU Parque Científico, BCMaterials, Bizkaia, Leioa E-48940, Spain.
- Department of Chemistry ''Giacomo Ciamician'', Alma Mater Studiorum - University of Bologna, Via Gobetti 85, Bologna 40129, Italy.
- IKERBASQUE, Basque Foundation for Science, Bilbao 48009, Spain.
ORCID-Profile: Stephania Rodríguez Muiña, Giovanni Valenti, Francisco Javier Del Campo
Förderung
- HORIZON EUROPE Marie Sklodowska-Curie Actions
- NextGenerationEU
Zitieren
APA
Rodríguez Muiña, S., Giagu, G., Valenti, G., & Del Campo, F. J. (2026). Versatile Electrochemiluminescence Imaging Platform for Multiplexed Biosensing and Point-of-Care Application. Chemical & Biomedical Imaging, 4(7), 1491-1500. https://doi.org/10.1021/cbmi.5c00274
Vancouver
Rodríguez Muiña S, Giagu G, Valenti G, Del Campo FJ. Versatile Electrochemiluminescence Imaging Platform for Multiplexed Biosensing and Point-of-Care Application. Chemical & Biomedical Imaging. 2026;4(7):1491-1500. doi:10.1021/cbmi.5c00274
BibTeX
@article{RodrguezMuia2026,
author = {Rodríguez Muiña, Stephania and Giagu, Gabriele and Valenti, Giovanni and Del Campo, Francisco Javier},
title = {Versatile Electrochemiluminescence Imaging Platform for Multiplexed Biosensing and Point-of-Care Application},
journal = {Chemical & Biomedical Imaging},
year = {2026},
volume = {4},
number = {7},
pages = {1491--1500},
doi = {10.1021/cbmi.5c00274}
}
EUVIMED