From Sensitivity to Temporal Fidelity in Wearable Sweat Sensing

Zeyu Lu, Marc Josep Montagut Marques, Kayo Hirose, Shinjiro Umezu iD

ACS Sensors · 12.11.2026

DOI: 10.1021/acssensors.6c04588 ↗

Kurz gefasst

Fachartikel in der Zeitschrift ACS Sensors, online erschienen am 12.11.2026. Verfasst von Zeyu Lu und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Modern Mechanical Engineering, Waseda University, 3-4-1 Okubo, Shinjuku-ku, Tokyo169-8555, Japan. Die Arbeit nennt 67 Literaturangaben. Gefördert von: Japan Science and Technology Agency.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Wearable sweat sensors are judged mainly by analytical sensitivity, yet a detection limit cannot tell when a measured concentration was secreted. At low or intermittent sweat rates, fluid retained in the device delays and mixes successive samples, so each reading reflects a distribution of sample ages rather than a single moment. In this Perspective, we argue that temporal fidelity should stand alongside sensitivity as a design target. We treat it as a property of the whole measurement chain, from biological transfer through sampling topology to sensor kinetics. Minimal transport and sensor models relate the tolerable transport time to the timescale of the physiological feature, and published sweat glucose and chloride traces test the resulting envelopes. Under the illustrative fidelity criteria used here, no transport threshold is universal. Hour-scale trends tolerate mean transport times on the order of 10 min, whereas features of 30 min or shorter generally require renewal within a few minutes or less. For a 30 min feature read by a 1 min sensor, the lowest reported resting sweat rate limits the coupled sensing volume over a 3 mm collection area to about 1 nL. Transport and sensor lags can also be indistinguishable in the device output, so each stage must be characterized separately. We therefore map five sampling architectures to the temporal claims each can support and propose a decision-first validation sequence with minimum reporting items. Sensitivity remains necessary, but interpretable sweat data also require sample age and renewal to be measured rather than assumed.

Schlagwörter

Stichwörter der Autor:innen

low-volume sensingmicrofluidicsresidence-time distributionsample agesensor kineticstemporal fidelitywearable sweat sensors

Autor:innen und Institutionen

  • Department of Modern Mechanical Engineering, Waseda University, 3-4-1 Okubo, Shinjuku-ku, Tokyo169-8555, Japan.
  • Department of Integrative Bioscience and Biomedical Engineering, Waseda University, 3-4-1 Okubo, Shinjuku-ku, Tokyo169-8555, Japan.
  • Department of Anesthesiology and Pain Relief Center, The University of Tokyo Hospital , 7-3-1 Hongo, Bunkyo-ku, Tokyo113-8655, Japan.

ORCID-Profile: Shinjiro Umezu

Förderung

  • Japan Science and Technology Agency

Zitieren

APA

Lu, Z., Josep Montagut Marques, M., Hirose, K., & Umezu, S. (2026). From Sensitivity to Temporal Fidelity in Wearable Sweat Sensing. ACS Sensors. https://doi.org/10.1021/acssensors.6c04588

Vancouver

Lu Z, Josep Montagut Marques M, Hirose K, Umezu S. From Sensitivity to Temporal Fidelity in Wearable Sweat Sensing. ACS Sensors. 2026. doi:10.1021/acssensors.6c04588

BibTeX
@article{Lu2026,
  author = {Lu, Zeyu and Josep Montagut Marques, Marc and Hirose, Kayo and Umezu, Shinjiro},
  title = {From Sensitivity to Temporal Fidelity in Wearable Sweat Sensing},
  journal = {ACS Sensors},
  year = {2026},
  doi = {10.1021/acssensors.6c04588}
}
Kostenlos starten →