EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Comparative Evaluation of Electrochemical and Fluorescent Readouts for Immunomagnetically Captured Waterborne Pathogens Using a Biochip with Bifunctional Magnetic Nanoparticles

Dharanivasan Gunasekaran iD, Abraham O Ogungbile iD, Izthak Icin iD, Sefi Vernick iD

ACS Measurement Science Au · 15.04.2026 · 6 (2) · 349-360

DOI: 10.1021/acsmeasuresciau.5c00161 ↗

Kurz gefasst

Fachartikel in der Zeitschrift ACS Measurement Science Au, online erschienen am 27.02.2026. Verfasst von Dharanivasan Gunasekaran und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Sensing, Information and Mechanization Engineering, Institute of Agricultural and Biosystems Engineering, Volcani Institute (ARO), Rish…. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Israel Water Authority.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Detection of waterborne pathogens benefits from measurement strategies that combine sensitivity with a practical workflow. We report a side-by-side metrological comparison of two orthogonal readouts measured on the same immunomagnetically captured cells using bifunctional Fe3O4 nanoparticles (MNPs) conjugated to anti-E. coli IgG and ferrocene (Fc). Under our protocol, these Fc-IgG MNPs achieved capturing efficiency up to 95% for Escherichia coli (E. coli) K12. The electrochemical readout, performed by a microfabricated chip (via differential pulse voltammetry of Fc) shows concentration-dependent signal suppression upon capture of E. coli K12, with an apparent detection limit as low as 10 cells·mL-1 and a broad dynamic range spanning 101 to 109 cells·mL-1. A complementary on-chip fluorescence readout (via Nile Red staining) provides visual corroboration and specificity support but exhibits reduced analytical sensitivity, approximately 4 orders of magnitude lower than the electrochemical approach, consistent with nanoparticle-induced quenching at lower cell counts. Across model and drinking-water matrices tested, the comparative performance trend is preserved, and the complete workflow can be completed in about 2 h from capture to readout. Taken together, these results present a comprehensive, internally consistent comparison on a single capture construct and clarify when electrochemistry versus fluorescence is fit-for-purpose in the context of immunomagnetic separation. The platform supports low-consumable testing with a reusable, unmodified electrochemical chip, is amenable to multiplexed array formats, and is readily compatible with portable potentiostats for on-site measurements.

Schlagwörter

Stichwörter der Autor:innen

bifunctional magnetic nanoparticlescomparative evaluationelectrochemical detectionimmunomagnetic separationmicrofabricated chipnile red fluorescenceorthogonal readouts

Autor:innen und Institutionen

  • Department of Sensing, Information and Mechanization Engineering, Institute of Agricultural and Biosystems Engineering, Volcani Institute (ARO), Rishon LeZion 5025001, Israel.
  • Department of Soil and Water Sciences, Robert H. Smith Faculty of Agriculture, Food and Environment, The Hebrew University of Jerusalem, P.O. Box 12, Rehovot 761001, Israel.
  • Department of Physical Electronics, Faculty of Engineering, Tel Aviv University, Tel Aviv 69978, Israel.

ORCID-Profile: Dharanivasan Gunasekaran, Abraham O Ogungbile, Izthak Icin, Sefi Vernick

Förderung

  • Israel Water Authority

Zitieren

APA

Gunasekaran, D., Ogungbile, A. O., Icin, I., & Vernick, S. (2026). Comparative Evaluation of Electrochemical and Fluorescent Readouts for Immunomagnetically Captured Waterborne Pathogens Using a Biochip with Bifunctional Magnetic Nanoparticles. ACS Measurement Science Au, 6(2), 349-360. https://doi.org/10.1021/acsmeasuresciau.5c00161

Vancouver

Gunasekaran D, Ogungbile AO, Icin I, Vernick S. Comparative Evaluation of Electrochemical and Fluorescent Readouts for Immunomagnetically Captured Waterborne Pathogens Using a Biochip with Bifunctional Magnetic Nanoparticles. ACS Measurement Science Au. 2026;6(2):349-360. doi:10.1021/acsmeasuresciau.5c00161

BibTeX
@article{Gunasekaran2026,
  author = {Gunasekaran, Dharanivasan and Ogungbile, Abraham O and Icin, Izthak and Vernick, Sefi},
  title = {Comparative Evaluation of Electrochemical and Fluorescent Readouts for Immunomagnetically Captured Waterborne Pathogens Using a Biochip with Bifunctional Magnetic Nanoparticles},
  journal = {ACS Measurement Science Au},
  year = {2026},
  volume = {6},
  number = {2},
  pages = {349--360},
  doi = {10.1021/acsmeasuresciau.5c00161}
}
Kostenlos starten →