SANDWiCH: An Oil-Free High-Throughput Microwell-Array Platform for Digital Loop-Mediated Isothermal Amplification

Jeonghwan Oh iD, Dong Hyun Han iD, Jongho Park iD, Juhyun Lee iD, Kyung Tae Noh iD, Je-Kyun Park iD

Analytical Chemistry · 10.11.2026 · 98 (44) · 30670-30681

DOI: 10.1021/acs.analchem.6c03950 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical Chemistry, online erschienen am 01.10.2026. Verfasst von Jeonghwan Oh und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology, Daejeon 34141, Republic of Korea. Hauptthemen laut MEDLINE: Nucleic Acid Amplification Techniques, High-Throughput Screening Assays, Molecular Diagnostic Techniques, DNA, Protozoan. Die Arbeit nennt 37 Literaturangaben. Gefördert von: National Research Foundation of Korea, Armed Forces Medical Research Institute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Digital nucleic acid amplification tests enable absolute molecular quantification by partitioning samples into independent reactors, but scalable digital loop-mediated isothermal amplification (LAMP) remains limited by oil-based compartmentalization, droplet stability requirements, and assembly-induced sample spreading in open microwell arrays. Here, we present serially-assembled nanodrop detection with a chip (SANDWiCH), an oil-free, high-throughput microwell-array digital LAMP platform designed for contamination-resistant multi-sample analysis. SANDWiCH integrates a 3 × 8 array layout, micropillar-assisted capillary loading, pressure-sensitive adhesive film sealing, and alignment-guided serial assembly to confine each sample to its assigned array before compartmentalization. This design enables spontaneous partitioning of reaction mixtures into 10,000 microwells, each with a volume of 7.85 pL, per array without pumps, actuators, or oil overlays, allowing 24 samples to be processed in parallel under identical amplification conditions. By varying the number of microwell-array units assigned per sample, the platform provides a tunable balance between assay throughput and quantification accuracy; the limit of quantification was reduced from 30.2 copies/μL in the highest-throughput mode to 1.3 copies/μL in the highest-accuracy mode. Using Plasmodium vivax as a target, quantitative analysis showed excellent agreement with theoretical predictions based on the Poisson distribution across six orders of magnitude (Pearson correlation coefficient, r = 0.991). Clinical validation with 60 blood-derived nucleic acid samples yielded 100% (95% CI, 88.4%−100%) sensitivity and 100% (95% CI, 88.4%−100%) specificity, while requiring only 0.29 μL of reagent per sample.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Nucleic Acid Amplification TechniquesHigh-Throughput Screening AssaysMolecular Diagnostic TechniquesDNA, ProtozoanPlasmodium falciparum

Stichwörter der Autor:innen

digital loop-mediated isothermal amplificationmalaria diagnosticsmicropillar-assisted fillingmicrowell chipoil-free compartmentalization

Substanzen

DNA, Protozoan

Autor:innen und Institutionen

  • Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology, Daejeon 34141, Republic of Korea.
  • Department of Infectious Disease Research, Armed Forces Medical Research Institute, Daejeon34059, Republic of Korea.

ORCID-Profile: Jeonghwan Oh, Dong Hyun Han, Jongho Park, Juhyun Lee, Kyung Tae Noh, Je-Kyun Park

Förderung

  • National Research Foundation of Korea
  • Armed Forces Medical Research Institute

Zitieren

APA

Oh, J., Han, D. H., Park, J., Lee, J., Noh, K. T., & Park, J. K. (2026). SANDWiCH: An Oil-Free High-Throughput Microwell-Array Platform for Digital Loop-Mediated Isothermal Amplification. Analytical Chemistry, 98(44), 30670-30681. https://doi.org/10.1021/acs.analchem.6c03950

Vancouver

Oh J, Han DH, Park J, Lee J, Noh KT, Park JK. SANDWiCH: An Oil-Free High-Throughput Microwell-Array Platform for Digital Loop-Mediated Isothermal Amplification. Analytical Chemistry. 2026;98(44):30670-30681. doi:10.1021/acs.analchem.6c03950

BibTeX
@article{Oh2026,
  author = {Oh, Jeonghwan and Han, Dong Hyun and Park, Jongho and Lee, Juhyun and Noh, Kyung Tae and Park, Je-Kyun},
  title = {SANDWiCH: An Oil-Free High-Throughput Microwell-Array Platform for Digital Loop-Mediated Isothermal Amplification},
  journal = {Analytical Chemistry},
  year = {2026},
  volume = {98},
  number = {44},
  pages = {30670--30681},
  doi = {10.1021/acs.analchem.6c03950}
}
Kostenlos starten →