EUVIMEDEUVIMED entdecken
Crossref

Ultrasensitive Quantification of Poloxamer 188 in Adeno-Associated Virus Products by All-Ion Fragmentation MS/MS

Sisi Zhang iD, Jing Wang, Hui Xiao iD, Ning Li

Analytical Chemistry · 10.11.2026 · 98 (44) · 30662-30669

DOI: 10.1021/acs.analchem.6c03755 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical Chemistry, online erschienen am 16.10.2026. Verfasst von Sisi Zhang und 3 weiteren Autor:innen. Beteiligte Einrichtung: Analytical and Biological Mass Spectrometry, Regeneron Pharmaceuticals Inc, 777 Old Saw Mill River Road, Tarrytown, New York10591-6706, United States. Hauptthemen laut MEDLINE: Poloxamer, Dependovirus, Tandem Mass Spectrometry. Die Arbeit nennt 18 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Poloxamer 188 (P188; Pluronic F68) is a nonionic surfactant frequently used in biotherapeutics, particularly in adeno-associated virus (AAV) formulations, in which it is present at trace concentrations (0.001%−0.005%). However, its accurate quantification is difficult, because of its low abundance, frequently limited sample volumes, and the ionization variability of its PEO−PPO−PEO triblock. Here, we developed an all-ion fragmentation (AIF) workflow with 44 Da and 58 Da neutral-loss-triggered data dependent (dd)-MS2 that fragments all precursors and performs quantification via the polyethylene glycol (PEG) diagnostic ion (m/z 133.1), thereby decreasing spectral complexity and adduct variability. In contrast to the most sensitive reported method, LC−MRM, using a 20 μL injection volume, which achieved a dynamic range of 250−10,000 ng/mL (0.000025%−0.001%), our method achieved a 0.00002% LLOQ with 1 μL injections, thus providing an ∼20-fold sensitivity gain that enables replicate analysis and a dynamic range 2 orders of magnitude wider (0.00002%−0.1%). Our assay is vendor-, formulation-, and LC-buffer-agnostic, and achieves ±10% accuracy and ∼10% CVs across AAV-relevant levels. Significant P188 losses attributable to container adsorption were observed; incorporation of 50% acetonitrile restored recovery and is recommended for sample preparation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PoloxamerDependovirusTandem Mass SpectrometryPolyethylene GlycolsLiquid Chromatography-Mass Spectrometry

Substanzen

PoloxamerPolyethylene Glycols

Autor:innen und Institutionen

  • Analytical and Biological Mass Spectrometry, Regeneron Pharmaceuticals Inc, 777 Old Saw Mill River Road, Tarrytown, New York10591-6706, United States.

ORCID-Profile: Sisi Zhang, Hui Xiao

Zitieren

APA

Zhang, S., Wang, J., Xiao, H., & Li, N. (2026). Ultrasensitive Quantification of Poloxamer 188 in Adeno-Associated Virus Products by All-Ion Fragmentation MS/MS. Analytical Chemistry, 98(44), 30662-30669. https://doi.org/10.1021/acs.analchem.6c03755

Vancouver

Zhang S, Wang J, Xiao H, Li N. Ultrasensitive Quantification of Poloxamer 188 in Adeno-Associated Virus Products by All-Ion Fragmentation MS/MS. Analytical Chemistry. 2026;98(44):30662-30669. doi:10.1021/acs.analchem.6c03755

BibTeX
@article{Zhang2026,
  author = {Zhang, Sisi and Wang, Jing and Xiao, Hui and Li, Ning},
  title = {Ultrasensitive Quantification of Poloxamer 188 in Adeno-Associated Virus Products by All-Ion Fragmentation MS/MS},
  journal = {Analytical Chemistry},
  year = {2026},
  volume = {98},
  number = {44},
  pages = {30662--30669},
  doi = {10.1021/acs.analchem.6c03755}
}
Kostenlos starten →