EUVIMEDEUVIMED entdecken
Crossref

Novel Ultrafast Sample-to-Answer Microfluidic Platform for Highly Sensitive On-Site Quantification of Avian Influenza Virus

Menghao Chai iD, Xiaofei Liu, Jingji Liu, Feng Chen, Qiongzhang Wang, Zhengying Si, Shanshan Yu, Yichao Gan, Liping Guo, Yude Yu, Yiqiang Fan iD, Songyin Qiu und 2 weitere

Analytical Chemistry · 2026

DOI: 10.1021/acs.analchem.5c06140 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical Chemistry, 2026. Verfasst von Menghao Chai und 13 weiteren Autor:innen. Beteiligte Einrichtung: School of Mechanical and Electrical Engineering, Beijing University of Chemical Technology, Beijing 100029, China. Hauptthemen laut MEDLINE: Influenza A Virus, H9N2 Subtype, Microfluidic Analytical Techniques, Influenza in Birds, Lab-On-A-Chip Devices. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 36 Literaturangaben. Gefördert von: National Key Research and Development Program of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Avian influenza virus (AIV) poses a global biosecurity threat, highlighting the need for rapid, accurate, and adaptable diagnostic tools to address its genetic diversity and complex transmission environments. Here, we report the development of FIGHTER, a novel ultrafast sample-to-answer microfluidic platform for the highly sensitive on-site detection of H9 and N2 genes to diagnose the H9N2 AIV subtype. The portable FIGHTER platform integrates an all-in-one plastic-aluminum membrane hybrid microfluidic chip with a dual-channel real-time PCR instrument, enabling on-site nucleic acid extraction in 5 min and reverse transcription-quantitative polymerase chain reaction with ultrafast thermal cycling in 18 min. In addition, the limit of detection of the platform is 1 copy/μL for both H9 and N2. The microfluidic chip with preloaded reagents supports simplified manual operation, and the battery-powered instrument features a user-friendly interface, allowing nonspecialists to operate it in resource-limited settings. Moreover, the validation with 60 field clinical samples showed a clinical sensitivity of 97.92% for H9 and 100% for N2, with 100% specificity for both H9 and N2. Hence, FIGHTER offers a simple, affordable, and rapid method for pathogen detection, particularly in complex environments where the conventional laboratory infrastructure is inaccessible.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Influenza A Virus, H9N2 SubtypeMicrofluidic Analytical TechniquesInfluenza in BirdsLab-On-A-Chip DevicesAnimalsLimit of DetectionBirdsRapid Diagnostic TestsReal-Time Polymerase Chain Reaction

Autor:innen und Institutionen

  • School of Mechanical and Electrical Engineering, Beijing University of Chemical Technology, Beijing 100029, China.
  • State Key Laboratory of Optoelectronic Materials and Devices, Institute of Semiconductors, Chinese Academy of Sciences, Beijing 100083, China.
  • Chinese Academy of Quality and Inspection &Testing, Beijing 100176, China.
  • School of Electrical and Electronic Engineering, North China Electric Power University, Beijing 102206, China.
  • School of Intelligent Manufacturing, Nanjing University of Science and Technology, Nanjing 214443, China.
  • College of Materials Science and Opto-Electronic Technology, University of Chinese Academy of Sciences, Beijing, 100049, China.

ORCID-Profile: Menghao Chai, Yiqiang Fan

Förderung

  • National Key Research and Development Program of China

Zitieren

APA

Chai, M., Liu, X., Liu, J., Chen, F., Wang, Q., Si, Z., Yu, S., Gan, Y., Guo, L., Yu, Y., Fan, Y., Qiu, S., Zhang, Y., & Li, Z. (2026). Novel Ultrafast Sample-to-Answer Microfluidic Platform for Highly Sensitive On-Site Quantification of Avian Influenza Virus. Analytical Chemistry. https://doi.org/10.1021/acs.analchem.5c06140

Vancouver

Chai M, Liu X, Liu J, Chen F, Wang Q, Si Z, et al. Novel Ultrafast Sample-to-Answer Microfluidic Platform for Highly Sensitive On-Site Quantification of Avian Influenza Virus. Analytical Chemistry. 2026. doi:10.1021/acs.analchem.5c06140

BibTeX
@article{Chai2026,
  author = {Chai, Menghao and Liu, Xiaofei and Liu, Jingji and Chen, Feng and Wang, Qiongzhang and Si, Zhengying and Yu, Shanshan and Gan, Yichao and Guo, Liping and Yu, Yude and Fan, Yiqiang and Qiu, Songyin and Zhang, Yajun and Li, Zhao},
  title = {Novel Ultrafast Sample-to-Answer Microfluidic Platform for Highly Sensitive On-Site Quantification of Avian Influenza Virus},
  journal = {Analytical Chemistry},
  year = {2026},
  doi = {10.1021/acs.analchem.5c06140}
}
Kostenlos starten →