EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Challenging the Database: Day-of-Analysis Calibration and UF Modeling for Reliable RRF Use in Medical Device Chemical Characterization

Michael Rush iD, J David Ricker, Pramod Prasad Poudel, Nuwan Kothalawala, Cary Watterson, Dmitriy Pastarnak

Analytical Chemistry · 2025

DOI: 10.1021/acs.analchem.5c04247 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical Chemistry, 2025. Verfasst von Michael Rush und 5 weiteren Autor:innen. Beteiligte Einrichtung: Edwards Lifesciences, 12050 Lone Peak Parkway, Draper, Utah 84020, United States. Hauptthemen laut MEDLINE: Equipment and Supplies. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Edwards Lifesciences.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Accurate quantitation during chemical characterization─also referred to as extractables and leachables (E&L)─within a toxicological risk assessment for medical device biocompatibility hinges on the appropriate application of relative response factors (RRFs). This study investigates the variability of RRFs across a chemically diverse set of compounds and evaluates the implications of quantitation model selection on analytical outcomes. Using gas chromatography-mass spectrometry (GC-MS) and liquid chromatography-mass spectrometry (LC-MS), we demonstrate that RRFs are highly context-dependent and influenced by factors such as ionization mode, compound class, concentration, and instrument conditions. We propose an approach to RRF determination, emphasizing day-of-analysis calibration and uncertainty factor modeling to improve reproducibility. Our findings support a statistically grounded application of analytical evaluation thresholds, enhancing the reliability of semiquantitative assessments in medical device biocompatibility.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Equipment and SuppliesCalibrationGas Chromatography-Mass SpectrometryChromatography, LiquidReproducibility of ResultsMass Spectrometry

Autor:innen und Institutionen

  • Edwards Lifesciences, 12050 Lone Peak Parkway, Draper, Utah 84020, United States.

ORCID-Profile: Michael Rush

Förderung

  • Edwards Lifesciences

Zitieren

APA

Rush, M., Ricker, J. D., Poudel, P. P., Kothalawala, N., Watterson, C., & Pastarnak, D. (2025). Challenging the Database: Day-of-Analysis Calibration and UF Modeling for Reliable RRF Use in Medical Device Chemical Characterization. Analytical Chemistry. https://doi.org/10.1021/acs.analchem.5c04247

Vancouver

Rush M, Ricker JD, Poudel PP, Kothalawala N, Watterson C, Pastarnak D. Challenging the Database: Day-of-Analysis Calibration and UF Modeling for Reliable RRF Use in Medical Device Chemical Characterization. Analytical Chemistry. 2025. doi:10.1021/acs.analchem.5c04247

BibTeX
@article{Rush2025,
  author = {Rush, Michael and Ricker, J David and Poudel, Pramod Prasad and Kothalawala, Nuwan and Watterson, Cary and Pastarnak, Dmitriy},
  title = {Challenging the Database: Day-of-Analysis Calibration and UF Modeling for Reliable RRF Use in Medical Device Chemical Characterization},
  journal = {Analytical Chemistry},
  year = {2025},
  doi = {10.1021/acs.analchem.5c04247}
}
Kostenlos starten →