EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Improving Spectral Similarity and Molecular Network Reliability through Noise Signal Filtering in MS/MS Spectra

Nicola Dalla Valle, Mar Garcia-Aloy iD, Peter Robatscher, Pietro Franceschi iD, Michael Oberhuber iD

Analytical Chemistry · 29.07.2025 · 97 (29) · 15873-15882

DOI: 10.1021/acs.analchem.5c02109 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical Chemistry, online erschienen am 17.07.2025. Verfasst von Nicola Dalla Valle und 4 weiteren Autor:innen. Beteiligte Einrichtung: University of Trento, Trento, TN 38100, Italy. Hauptthemen laut MEDLINE: Tandem Mass Spectrometry, Metabolomics. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In mass spectrometry, fragmentation spectra play a central role in compound identification. However, noise in MS/MS spectra can significantly impact similarity scores and molecular network (MN) reliability, leading to inaccurate compound annotation in untargeted metabolomics. This work investigates the influence of noise on MS/MS similarity scores and molecular network structure. Noise elimination increased similarity scores for homologous spectra, enhancing match affordability. In MNs, effective noise management improved network structure, resulting in more interpretable networks with fewer edges and enhanced clustering, decreasing false-positive connections. To quantitatively assess these improvements, a minimum spanning tree (MST) analysis was performed, revealing denser regions in the denoised MNs. An increasing cutoff of noise threshold can lead to an overlay between two or more different compound spectra. A data-specific workflow was developed to identify the optimal threshold for denoising, balancing spectra quality and network integrity during noise elimination, by incorporating statistics calculated on the distribution of the MST distances and the number of fragment ions, which could be explained by an in-silico fragmentation algorithm. Finally, a faster-tailored denoising method, based solely on the intensity of individual spectral ions, demonstrated performance comparable to the previously cited fixed threshold approaches.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tandem Mass SpectrometryMetabolomicsAlgorithmsSignal-To-Noise Ratio

Autor:innen und Institutionen

  • University of Trento, Trento, TN 38100, Italy.
  • Research Innovation Centre, Fondazione Edmund Mach, San Michele a/A, TN 38010, Italy.
  • Laimburg Research Centre, Laimburg 6 - Pfatten (Vadena), Auer (Ora), BZ 39040, Italy.

ORCID-Profile: Mar Garcia-Aloy, Pietro Franceschi, Michael Oberhuber

Zitieren

APA

Dalla Valle, N., Garcia-Aloy, M., Robatscher, P., Franceschi, P., & Oberhuber, M. (2025). Improving Spectral Similarity and Molecular Network Reliability through Noise Signal Filtering in MS/MS Spectra. Analytical Chemistry, 97(29), 15873-15882. https://doi.org/10.1021/acs.analchem.5c02109

Vancouver

Dalla Valle N, Garcia-Aloy M, Robatscher P, Franceschi P, Oberhuber M. Improving Spectral Similarity and Molecular Network Reliability through Noise Signal Filtering in MS/MS Spectra. Analytical Chemistry. 2025;97(29):15873-15882. doi:10.1021/acs.analchem.5c02109

BibTeX
@article{DallaValle2025,
  author = {Dalla Valle, Nicola and Garcia-Aloy, Mar and Robatscher, Peter and Franceschi, Pietro and Oberhuber, Michael},
  title = {Improving Spectral Similarity and Molecular Network Reliability through Noise Signal Filtering in MS/MS Spectra},
  journal = {Analytical Chemistry},
  year = {2025},
  volume = {97},
  number = {29},
  pages = {15873--15882},
  doi = {10.1021/acs.analchem.5c02109}
}
Kostenlos starten →