EUVIMEDEUVIMED entdecken
Crossref

Fluorescence Aptasensor for sST2 Detection Using In Vitro Selected Aptamers

Shuqian Qiu, Li Chen, Dongqing Zhuang, Yue Cao, Renli Wei, Xiaoai Cao, Yongshou Chen, Xiaofeng Lai, Shuiliang Wang, Yue Lin, Zhenyu Lin iD, Shenghang Zhang

Analytical Chemistry · 27.05.2025 · 97 (20) · 10910-10918

DOI: 10.1021/acs.analchem.5c01855 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical Chemistry, online erschienen am 15.05.2025. Verfasst von Shuqian Qiu und 11 weiteren Autor:innen. Beteiligte Einrichtung: Fujian Key Laboratory of Aptamers Technology, Fuzong Teaching Hospital of Fujian University of Traditional Chinese Medicine (900th Hospital), Fuzhou …. Hauptthemen laut MEDLINE: Aptamers, Nucleotide, Interleukin-1 Receptor-Like 1 Protein, Biosensing Techniques. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 32 Literaturangaben. Gefördert von: Fujian Medical University, 900 Hospital of the People's liberation Army Joint Logistic Support Force, National filing science and technology project.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Soluble suppression of tumorigenicity 2 (sST2) is a critical biomarker for heart failure (HF) diagnosis and prognosis, yet conventional antibody-based detection methods suffer from time-consuming protocols and high costs and involve complex detection procedures. To address these challenges, we first screened high-affinity aptamers under clinically relevant conditions and then coupled with the CRISPR/Cas12a system to develop a fluorescence aptasensor for rapid and sensitive sST2 detection. A serum matrix was introduced during aptamer selection to enhance specificity and anti-interference performance in real biological environments. Three sST2-specific aptamers (Apt-1, Apt-2, and Apt-3) were identified with dissociation constants ( K D ) of 8.42, 46.08, and 25.02 nM, respectively. Among these, Apt-1 demonstrated superior performance, which was utilized to construct a fluorescence biosensor combining aptamer recognition with CRISPR/Cas12a trans-cleavage signal amplification. The sensor achieved a broad linear detection range (5–120 ng/mL) and an ultralow limit of detection (LOD, 0.816 ng/mL) when applied in detecting sST2 in both the buffer and human serum. Notably, the platform exhibited exceptional resistance to interference from HF-related proteins and maintained high accuracy in clinical serum samples, showing a strong correlation ( R 2 = 0.9794) with enzyme-linked immunosorbent assay (ELISA) results. By integration of serum-matrix screening and CRISPR-based signal enhancement, this work establishes a robust, cost-effective, and rapid diagnostic tool for sST2 detection.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Aptamers, NucleotideInterleukin-1 Receptor-Like 1 ProteinBiosensing TechniquesHumansLimit of DetectionFluorescenceCRISPR-Cas Systems

Substanzen

Aptamers, NucleotideInterleukin-1 Receptor-Like 1 ProteinIL1RL1 protein, human

Autor:innen und Institutionen

  • Fujian Key Laboratory of Aptamers Technology, Fuzong Teaching Hospital of Fujian University of Traditional Chinese Medicine (900th Hospital), Fuzhou 350025, China.
  • Department of Clinical Medicine, Fuzong Clinical Medical College of Fujian Medical University, Fuzhou 350025, China.
  • Department of Clinical Laboratory Medicine, Dongfang Hospital of School of Medicine, Xiamen University, Fuzhou 350025, China.
  • Ministry of Education Key Laboratory for Analytical Science of Food Safety and Biology, Fujian Provincial Key Laboratory of Analysis and Detection for Food Safety, College of Chemistry, Fuzhou University, Fuzhou, Fujian…

ORCID-Profile: Zhenyu Lin

Förderung

  • Fujian Medical University
  • 900 Hospital of the People's liberation Army Joint Logistic Support Force
  • National filing science and technology project
  • Joint Funds for the innovation of science and Technology, Fujian province
  • Fujian Clinical Research Center for Aptamer-based Precision Testing

Zitieren

APA

Qiu, S., Chen, L., Zhuang, D., Cao, Y., Wei, R., Cao, X., Chen, Y., Lai, X., Wang, S., Lin, Y., Lin, Z., & Zhang, S. (2025). Fluorescence Aptasensor for sST2 Detection Using In Vitro Selected Aptamers. Analytical Chemistry, 97(20), 10910-10918. https://doi.org/10.1021/acs.analchem.5c01855

Vancouver

Qiu S, Chen L, Zhuang D, Cao Y, Wei R, Cao X, et al. Fluorescence Aptasensor for sST2 Detection Using In Vitro Selected Aptamers. Analytical Chemistry. 2025;97(20):10910-10918. doi:10.1021/acs.analchem.5c01855

BibTeX
@article{Qiu2025,
  author = {Qiu, Shuqian and Chen, Li and Zhuang, Dongqing and Cao, Yue and Wei, Renli and Cao, Xiaoai and Chen, Yongshou and Lai, Xiaofeng and Wang, Shuiliang and Lin, Yue and Lin, Zhenyu and Zhang, Shenghang},
  title = {Fluorescence Aptasensor for sST2 Detection Using In Vitro Selected Aptamers},
  journal = {Analytical Chemistry},
  year = {2025},
  volume = {97},
  number = {20},
  pages = {10910--10918},
  doi = {10.1021/acs.analchem.5c01855}
}