EUVIMEDEUVIMED entdecken
Crossref

On-Chip Isolation and Reciprocal Signal Amplification Detection of Tumor-Derived Exosomes in Dual-Control Microfluidic Device

Junhe Ma, Kexin Li, Zhaofan Duan, Xuexin Yang, Guodong Zhou, Sujuan Ye iD

Analytical Chemistry · 08.04.2025 · 97 (13) · 7483-7489

DOI: 10.1021/acs.analchem.5c00426 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical Chemistry, online erschienen am 28.03.2025. Verfasst von Junhe Ma und 5 weiteren Autor:innen. Beteiligte Einrichtung: Key Laboratory of Optic-electric Sensing and Analytical Chemistry for Life Science, MOE; College of Chemistry and Molecular Engineering, Qingdao Univ…. Hauptthemen laut MEDLINE: Exosomes, Lab-On-A-Chip Devices, Microfluidic Analytical Techniques, Neoplasms. Die Arbeit wurde bisher 18-mal zitiert (Crossref) und nennt 31 Literaturangaben. Gefördert von: National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The detection of exosomes is critical for health monitoring and disease diagnosis. However, their small size and low concentration present significant challenges. In this study, we designed a dual-control microchip integrated with a surface-enhanced Raman scattering (SERS) signal amplification detection method. By employing separate chambers for isolation and detection, this method achieves magnetic separation control and DNA cascade signal amplification with electrokinetic enrichment detection. The magnetic separation step captures and isolates exosomes in a magnetic-controlled reaction chamber, releasing a signal-switching strand that translates exosome recognition into a DNA signal amplification process. The DNA cascade reciprocal signal amplification reaction is performed in an electrokinetic enrichment reaction chamber, significantly improving detection efficiency and signal intensity. In addition, absolute-value coupled data processing reduces background interference. These unique merits enable precise and highly efficient assay of exosomes. This dual-control microchip signal amplification sensor exhibits remarkable sensitivity, rapid detection times, with a detection limit of 10.9 particles/μL and a reaction time of 35 min, and successful application to real sample analysis. The platform offers a viable, accurate, and portable solution for medical point-of-care testing.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ExosomesLab-On-A-Chip DevicesMicrofluidic Analytical TechniquesNeoplasmsHumansSpectrum Analysis, RamanNucleic Acid Amplification TechniquesLimit of DetectionDNA

Substanzen

DNA

Autor:innen und Institutionen

  • Key Laboratory of Optic-electric Sensing and Analytical Chemistry for Life Science, MOE; College of Chemistry and Molecular Engineering, Qingdao University of Science and Technology, Qingdao 266042, P. R. China.

ORCID-Profile: Sujuan Ye

Förderung

  • National Natural Science Foundation of China

Zitieren

APA

Ma, J., Li, K., Duan, Z., Yang, X., Zhou, G., & Ye, S. (2025). On-Chip Isolation and Reciprocal Signal Amplification Detection of Tumor-Derived Exosomes in Dual-Control Microfluidic Device. Analytical Chemistry, 97(13), 7483-7489. https://doi.org/10.1021/acs.analchem.5c00426

Vancouver

Ma J, Li K, Duan Z, Yang X, Zhou G, Ye S. On-Chip Isolation and Reciprocal Signal Amplification Detection of Tumor-Derived Exosomes in Dual-Control Microfluidic Device. Analytical Chemistry. 2025;97(13):7483-7489. doi:10.1021/acs.analchem.5c00426

BibTeX
@article{Ma2025,
  author = {Ma, Junhe and Li, Kexin and Duan, Zhaofan and Yang, Xuexin and Zhou, Guodong and Ye, Sujuan},
  title = {On-Chip Isolation and Reciprocal Signal Amplification Detection of Tumor-Derived Exosomes in Dual-Control Microfluidic Device},
  journal = {Analytical Chemistry},
  year = {2025},
  volume = {97},
  number = {13},
  pages = {7483--7489},
  doi = {10.1021/acs.analchem.5c00426}
}