MobiLipid: A Tool for Enhancing CCS Quality Control of Ion Mobility–Mass Spectrometry Lipidomics by Internal Standardization
Analytical Chemistry · 2024
DOI: 10.1021/acs.analchem.4c01253 ↗Kurz gefasst
Fachartikel in der Zeitschrift Analytical Chemistry, 2024. Verfasst von Felina Hildebrand und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Analytical Chemistry, Faculty of Chemistry, University of Vienna, Waehringer Str. 38, 1090 Vienna, Austria. Hauptthemen laut MEDLINE: Lipidomics, Lipids, Mass Spectrometry. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Austrian Science Fund.
Themen: Mobilisation
Abstract (Originalsprache)
Ion mobility-mass spectrometry (IM-MS) offers benefits for lipidomics by obtaining IM-derived collision cross sections (CCS), a conditional property of an ion that can enhance lipid identification. While drift tube (DT) IM-MS retains a direct link to the primary experimental method to derive CCS values, other IM technologies rely solely on external CCS calibration, posing challenges due to dissimilar chemical properties between lipids and calibrants. To address this, we introduce MobiLipid, a novel tool facilitating the CCS quality control of IM-MS lipidomics workflows by internal standardization. MobiLipid utilizes a newly established DTCCSN2 library for uniformly (U)13C-labeled lipids, derived from a U13C-labeled yeast extract, containing 377 DTCCSN2 values. This automated open-source R Markdown tool enables internal monitoring and straightforward compensation for CCSN2 biases. It supports lipid class- and adduct-specific CCS corrections, requiring only three U13C-labeled lipids per lipid class-adduct combination across 10 lipid classes without requiring additional external measurements. The applicability of MobiLipid is demonstrated for trapped IM (TIM)-MS measurements of an unlabeled yeast extract spiked with U13C-labeled lipids. Monitoring the CCSN2 biases of TIMCCSN2 values compared to DTCCSN2 library entries utilizing MobiLipid resulted in mean absolute biases of 0.78% and 0.33% in positive and negative ionization mode, respectively. By applying the CCS correction integrated into the tool for the exemplary data set, the mean absolute CCSN2 biases of 10 lipid classes could be reduced to approximately 0%.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Analytical Chemistry, Faculty of Chemistry, University of Vienna, Waehringer Str. 38, 1090 Vienna, Austria.
- Vienna Doctoral School in Chemistry (DoSChem), University of Vienna, Waehringer Str. 42, 1090 Vienna, Austria.
- Vienna Metabolomics Center (VIME), University of Vienna, Althanstr. 14, 1090 Vienna, Austria.
- BOKU University, Department of Chemistry, Institute of Analytical Chemistry, Muthgasse 18, 1190 Vienna, Austria.
ORCID-Profile: Felina Hildebrand, Gunda Koellensperger, Tim Causon
Förderung
- Austrian Science Fund
Zitieren
APA
Hildebrand, F., Koellensperger, G., & Causon, T. (2024). MobiLipid: A Tool for Enhancing CCS Quality Control of Ion Mobility–Mass Spectrometry Lipidomics by Internal Standardization. Analytical Chemistry. https://doi.org/10.1021/acs.analchem.4c01253
Vancouver
Hildebrand F, Koellensperger G, Causon T. MobiLipid: A Tool for Enhancing CCS Quality Control of Ion Mobility–Mass Spectrometry Lipidomics by Internal Standardization. Analytical Chemistry. 2024. doi:10.1021/acs.analchem.4c01253
BibTeX
@article{Hildebrand2024,
author = {Hildebrand, Felina and Koellensperger, Gunda and Causon, Tim},
title = {MobiLipid: A Tool for Enhancing CCS Quality Control of Ion Mobility–Mass Spectrometry Lipidomics by Internal Standardization},
journal = {Analytical Chemistry},
year = {2024},
doi = {10.1021/acs.analchem.4c01253}
}
EUVIMED