Two-Dimensional Multi-parameter Cytometry Platform for Single-Cell Analysis

Chengxin Wu, Xue Men, Meijun Liu, Yujia Wei, Xing Wei, Yong-Liang Yu iD, Zhang-Run Xu iD, Ming-Li Chen iD, Jian-Hua Wang iD

Analytical Chemistry · 2023

DOI: 10.1021/acs.analchem.3c02457 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Analytical Chemistry, 2023. Verfasst von Chengxin Wu und 8 weiteren Autor:innen. Beteiligte Einrichtung: Research Center for Analytical Sciences, Department of Chemistry, College of Sciences, Northeastern University, Shenyang 110819, China. Hauptthemen laut MEDLINE: Research Design, Algorithms. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 29 Literaturangaben. Gefördert von: Fundamental Research Funds for the Central Universities, National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A 2D flow cytometry platform, known as CytoLM Plus, was developed for multi-parameter single-cell analysis. Single particles or cells after hydrodynamic alignment in a microfluidic unit undergo first-dimension fluorescence and side scattering dual-channel optical detection. They were thereafter immediately directed to ICP-MS by connecting the microfluidic unit with a high-efficiency nebulizer to facilitate the second-dimension ICP-MS detection. Flow cytometry measurements of fluorescent microspheres evaluated the performance of CytoLM Plus for optical detection. 6434 fluorescence bursts were observed with a valid signal proportion as high as 99.7%. After signal unification and gating analysis, 6067 sets of single-particle signals were obtained with 6.6 and 6.2% deviations for fluorescence burst area and height, respectively. This is fairly comparable with that achieved by a commercial flow cytometer. Afterward, CytoLM Plus was evaluated by 2D flow cytometry measurement of Ag+-incubated and AO-stained MCF-7 cells. A program for 2D single-cell signal unification was developed based on the algorithm of screening in lag time window. In the present case, a lag time window of -4.2 ± 0.09 s was determined by cross-correlation analysis and two-parameter optimization, which efficiently unified the concurrent single-cell signals from fluorescence, side scattering, and ICP-MS. A total of 495 sets of concurrent 2D signals were screened out, and the statistical analysis of these single-cell signals ensured 2D multi-parameter single-cell analysis and data elucidation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Research DesignAlgorithmsHumansColoring AgentsFlow CytometrySingle-Cell Analysis

Substanzen

Coloring Agents

Autor:innen und Institutionen

  • Research Center for Analytical Sciences, Department of Chemistry, College of Sciences, Northeastern University, Shenyang 110819, China.
  • School of Chemistry and Environment, Yunnan Minzu University, Kunming 650504, China.

ORCID-Profile: Yong-Liang Yu, Zhang-Run Xu, Ming-Li Chen, Jian-Hua Wang

Förderung

  • Fundamental Research Funds for the Central Universities
  • National Natural Science Foundation of China

Zitieren

APA

Wu, C., Men, X., Liu, M., Wei, Y., Wei, X., Yu, Y. L., Xu, Z. R., Chen, M. L., & Wang, J. H. (2023). Two-Dimensional Multi-parameter Cytometry Platform for Single-Cell Analysis. Analytical Chemistry. https://doi.org/10.1021/acs.analchem.3c02457

Vancouver

Wu C, Men X, Liu M, Wei Y, Wei X, Yu YL, et al. Two-Dimensional Multi-parameter Cytometry Platform for Single-Cell Analysis. Analytical Chemistry. 2023. doi:10.1021/acs.analchem.3c02457

BibTeX
@article{Wu2023,
  author = {Wu, Chengxin and Men, Xue and Liu, Meijun and Wei, Yujia and Wei, Xing and Yu, Yong-Liang and Xu, Zhang-Run and Chen, Ming-Li and Wang, Jian-Hua},
  title = {Two-Dimensional Multi-parameter Cytometry Platform for Single-Cell Analysis},
  journal = {Analytical Chemistry},
  year = {2023},
  doi = {10.1021/acs.analchem.3c02457}
}
Kostenlos starten →