A Chemiluminescence Sensor for the Detection of α-Fetoprotein and Carcinoembryonic Antigen Based on Dual-Aptamer Functionalized Magnetic Silicon Composite
Analytical Chemistry · 2023
DOI: 10.1021/acs.analchem.3c01056 ↗Kurz gefasst
Fachartikel in der Zeitschrift Analytical Chemistry, 2023. Verfasst von Yuanling Sun und 4 weiteren Autor:innen. Beteiligte Einrichtung: Key Laboratory of Interfacial Reaction & Sensing Analysis in Universities of Shandong, School of Chemistry and Chemical Engineering, University of Ji…. Hauptthemen laut MEDLINE: DNA, Catalytic, Aptamers, Nucleotide, Metal Nanoparticles, Biosensing Techniques. Die Arbeit wurde bisher 48-mal zitiert (Crossref) und nennt 35 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Natural Science Foundation of Shandong Province, Horizontal Scientific Research Project of China.
Abstract (Originalsprache)
In this work, a dual-aptamer functionalized magnetic silicon composite was prepared and used to construct a chemiluminescence (CL) sensor for the detection of α-fetoprotein (AFP) and carcinoembryonic antigen (CEA). First, SiO2@Fe3O4 was prepared, and polydiallyl dimethylammonium chloride (PDDA) and AuNPs were sequentially loaded on SiO2@Fe3O4. Subsequently, the complementary strand of CEA aptamer (cDNA2) and the aptamer of AFP (Apt1) were attached to AuNPs/PDDA-SiO2@Fe3O4. Then, the aptamer of CEA (Apt2) and G quadruplex peroxide-mimicking enzyme (G-DNAzyme) were sequentially connected to cDNA2, leading to the final composite. Then, the composite was used to construct a CL sensor. When AFP is present, it will combine with Apt1 on the composite to hinder the catalytic ability of AuNPs to luminol-H2O2, achieving AFP detection. When CEA is present, it will recognize and bind with Apt2, so G-DNAzyme is released to solution and catalyzes the reaction of luminol-H2O2 to achieve CEA determination. After the application of the prepared composite, AFP and CEA were detected in the magnetic medium and supernatant, respectively, after simple magnetic separation. Therefore, the detection of multiple liver cancer markers is realized through the CL technology without additional instruments or technology, which broadens the application range of CL technology. The sensor for detecting AFP and CEA shows wide linear ranges of 1.0 × 10-4 to 1.0 ng·mL-1 and 0.0001-0.5 ng·mL-1 and low detection limits of 6.7 × 10-5 ng·mL-1 and 3.2 × 10-5 ng·mL-1, respectively. Finally, the sensor was successfully used to detect CEA and AFP in serum samples and provides great potential for detection of multiple liver cancer markers in early clinical diagnosis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Key Laboratory of Interfacial Reaction & Sensing Analysis in Universities of Shandong, School of Chemistry and Chemical Engineering, University of Jinan, Jinan 250022, PR China.
ORCID-Profile: Chuannan Luo
Förderung
- National Natural Science Foundation of China
- Natural Science Foundation of Shandong Province
- Horizontal Scientific Research Project of China
Zitieren
APA
Sun, Y., Hou, Y., Cao, T., Luo, C., & Wei, Q. (2023). A Chemiluminescence Sensor for the Detection of α-Fetoprotein and Carcinoembryonic Antigen Based on Dual-Aptamer Functionalized Magnetic Silicon Composite. Analytical Chemistry. https://doi.org/10.1021/acs.analchem.3c01056
Vancouver
Sun Y, Hou Y, Cao T, Luo C, Wei Q. A Chemiluminescence Sensor for the Detection of α-Fetoprotein and Carcinoembryonic Antigen Based on Dual-Aptamer Functionalized Magnetic Silicon Composite. Analytical Chemistry. 2023. doi:10.1021/acs.analchem.3c01056
BibTeX
@article{Sun2023,
author = {Sun, Yuanling and Hou, Yanan and Cao, Tianzi and Luo, Chuannan and Wei, Qin},
title = {A Chemiluminescence Sensor for the Detection of α-Fetoprotein and Carcinoembryonic Antigen Based on Dual-Aptamer Functionalized Magnetic Silicon Composite},
journal = {Analytical Chemistry},
year = {2023},
doi = {10.1021/acs.analchem.3c01056}
}