A preliminary study of the efficacy of bimodal versus unimodal stumble recovery responses in a powered knee prosthesis
Wearable Technologies · 2026 · 7
DOI: 10.1017/wtc.2026.10040 ↗Kurz gefasst
Fachartikel in der Zeitschrift Wearable Technologies, online erschienen am 07.04.2026. Verfasst von Shane T King und 4 weiteren Autor:innen. Beteiligte Einrichtung: Mechanical Engineering, Vanderbilt University, USA. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Eunice Kennedy Shriver National Institute of Child Health and Human Development.
Abstract (Originalsprache)
Transfemoral prosthesis users demonstrate a higher fall rate due to tripping than able-bodied controls in previous laboratory studies. In particular, early swing demonstrates the greatest disparity, where able-bodied controls typically utilize an elevating strategy to cross the obstacle in the same stride that the perturbation occurs, rather than the lowering strategy, where swing is ended prematurely and the obstacle is crossed in the following stride. However, due to the passive nature of most commercial knee prostheses, the elevating strategy is largely inaccessible to prosthesis users, potentially contributing to the increased fall rate in early swing. To investigate the effects of reintroducing the elevating strategy to transfemoral prosthesis users, a bimodal stumble recovery controller was developed for a powered knee prosthesis that utilized the elevating and lowering recovery strategies, selected based on the post-impact kinematics of the prosthesis. The Bimodal controller was compared to a unimodal controller that only used the lowering strategy. Three transfemoral prosthesis users underwent a series of treadmill-based obstacle perturbations with each controller following an acclimation period. All participants successfully used the elevating response in the early swing phase. On average, the elevating response reduced the disturbance to participants’ trunk kinematics and the reliance on harness support. While the Bimodal controller sometimes resulted in a recovery strategy mismatch for two participants, the mismatch still resulted in outcome metrics comparable to the unimodal controller. Overall, results suggest that the inclusion of the elevating and lowering strategies may improve stumble recovery outcomes for some transfemoral prosthesis users.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Mechanical Engineering, Vanderbilt University, USA.
- Electrical Engineering, Vanderbilt University, USA.
- Biomedical Engineering, Vanderbilt University, USA.
- Physical Medicine & Rehabilitation, Vanderbilt University, USA.
ORCID-Profile: Shane T King
Förderung
- Eunice Kennedy Shriver National Institute of Child Health and Human Development
Zitieren
APA
King, S. T., Eveld, M. E., Vailati, L. G., Zelik, K. E., & Goldfarb, M. (2026). A preliminary study of the efficacy of bimodal versus unimodal stumble recovery responses in a powered knee prosthesis. Wearable Technologies, 7. https://doi.org/10.1017/wtc.2026.10040
Vancouver
King ST, Eveld ME, Vailati LG, Zelik KE, Goldfarb M. A preliminary study of the efficacy of bimodal versus unimodal stumble recovery responses in a powered knee prosthesis. Wearable Technologies. 2026;7. doi:10.1017/wtc.2026.10040
BibTeX
@article{King2026,
author = {King, Shane T and Eveld, Maura E and Vailati, Leo G and Zelik, Karl E and Goldfarb, Michael},
title = {A preliminary study of the efficacy of bimodal versus unimodal stumble recovery responses in a powered knee prosthesis},
journal = {Wearable Technologies},
year = {2026},
volume = {7},
doi = {10.1017/wtc.2026.10040}
}