Patent ductus arteriosus closure in small infants weighing less than 1500 g using KA micro plug: steps and tips
Cardiology in the Young · 2024
DOI: 10.1017/s1047951124025629 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cardiology in the Young, 2024. Verfasst von Bassel Mohammad Nijres und 5 weiteren Autor:innen. Beteiligte Einrichtung: Pediatric Cardiology Section, Department of Pediatrics, King Faisal Specialist Hospital and Research Center, Jeddah, Saudi Arabia. Hauptthemen laut MEDLINE: Ductus Arteriosus, Patent, Cardiac Catheterization, Septal Occluder Device, Gestational Age. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben.
Abstract (Originalsprache)
Background
Transcatheter patent ductus arteriosus closure in premature infants is gaining attention with rapidly growing experience. The KA micro plug device poses many favourable features for ductus arteriosus closure in premature infants. The study aimed to describe the technique and outcomes of a KA micro plug in closing the ductus arteriosus in premature infants weighing less than 1500 g.
Methods
This is a single-centre retrospective study in premature infants with patent ductus arteriosus weighing less than 1500 g who underwent off-label use of a KA micro plug at the Stead Family Children’s Hospital from February 2022 to December 2023.
Results
Eighteen premature infants weighing less than 1500 g underwent attempted transcatheter ductus arteriosus closure using a KA micro plug device. Median weight, age, and corrected gestational age at the procedure were 943 g (682–1225), 26 days (9–79), and 28.5 weeks (25.6–32), respectively. The procedure was successful with complete closure achieved in all patients. No major haemodynamic instabilities were encountered. No catheterization-related complications were encountered. On a median follow-up of 10.9 months (0.1–19), all patients are alive with the device in a good position without residual shunt or abnormal findings.
Conclusion
KA micro plug device for patent ductus arteriosus in infants weighing less than 1500 g is feasible, safe, and effective. We recommend that manufacturers of devices used to close the ductus arteriosus in small infants enhance both the devices and their delivery systems to simplify the procedure and mitigate the risk of haemodynamic instability.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Pediatric Cardiology Section, Department of Pediatrics, King Faisal Specialist Hospital and Research Center, Jeddah, Saudi Arabia.
- Division of Pediatric Cardiology, Stead Family Children's Hospital, University of Iowa, Iowa City, IA, USA.
- Division of Pediatric Anesthesia, Stead Family Children's Hospital, University of Iowa, Iowa City, IA, USA.
- Division of Neonatology, Stead Family Children's Hospital, University of Iowa, Iowa City, IA, USA.
ORCID-Profile: Bassel Mohammad Nijres, Adrianne Rahde Bischoff
Zitieren
APA
Mohammad Nijres, B., Hassan, M., Windsor, J., Bischoff, A. R., McNamara, P. J., & Aldoss, O. (2024). Patent ductus arteriosus closure in small infants weighing less than 1500 g using KA micro plug: steps and tips. Cardiology in the Young. https://doi.org/10.1017/s1047951124025629
Vancouver
Mohammad Nijres B, Hassan M, Windsor J, Bischoff AR, McNamara PJ, Aldoss O. Patent ductus arteriosus closure in small infants weighing less than 1500 g using KA micro plug: steps and tips. Cardiology in the Young. 2024. doi:10.1017/s1047951124025629
BibTeX
@article{MohammadNijres2024,
author = {Mohammad Nijres, Bassel and Hassan, Moeez and Windsor, Jimmy and Bischoff, Adrianne Rahde and McNamara, Patrick J and Aldoss, Osamah},
title = {Patent ductus arteriosus closure in small infants weighing less than 1500 g using KA micro plug: steps and tips},
journal = {Cardiology in the Young},
year = {2024},
doi = {10.1017/s1047951124025629}
}
EUVIMED