Risk factors associated with device embolisation or malposition during transcatheter closure of patent ductus arteriosus
Cardiology in the Young · 2023-10 · 33 (10) · 2041-2048
DOI: 10.1017/s1047951122003973 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cardiology in the Young, online erschienen am 14.12.2022. Verfasst von Zeeshan A Mumtaz und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pediatric Cardiology, Institute of Cardiovascular Diseases, Madras Medical Mission, Chennai, India. Hauptthemen laut MEDLINE: Ductus Arteriosus, Patent, Embolization, Therapeutic, Septal Occluder Device. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 15 Literaturangaben.
Abstract (Originalsprache)
Background
Device embolisation is a serious adverse event during transcatheter duct closure. This study analyses risk factors for embolisation.
Methods
Demographic parameters, echocardiographic anatomy, haemodynamics, and procedural characteristics of consecutive duct closures in a tertiary centre over 8 years were analysed. Procedures complicated by embolisation were compared to uncomplicated procedures.
Results
Fifteen embolisations occurred during 376 procedures. All except one embolisation were in infants. The pulmonary artery: aortic pressure ratio was 0.78 ± 0.22. Embolisation was seen significantly more commonly in Type C tubular ducts. Vascular plugs were more significantly associated with embolisations. Logistic regression analysis showed device embolisation was significantly higher in age group of < 6 months compared to 6–12 months (p = 0.02), higher in those with tubular ducts versus conical ducts (p = 0.003), use of vascular plugs compared to conventional duct occluders (p = 0.05), and in duct closure with undersized devices (p = 0.001). There was no in-hospital mortality. Three patients needed surgical retrieval while others were successfully managed in catheterisation laboratory.
Conclusions
Device embolisation complicates 4% of transcatheter duct closures, with need for surgery in one-fifth of them. Larger ducts with high pulmonary artery pressures in younger and smaller infants are more often associated with device embolisation. Tubular ducts are more prone for embolisation compared to usual conical ducts. Softer vascular plugs are often associated with embolisations. Intentional device undersizing to avoid vascular obstruction in small patients is a frequent risk factor for embolisation. Precise echocardiographic measurements, correct occluder choice, proper technique and additional care in patients with high pulmonary artery pressures are mandatory to minimise embolisations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Pediatric Cardiology, Institute of Cardiovascular Diseases, Madras Medical Mission, Chennai, India.
ORCID-Profile: Zeeshan A Mumtaz, Pramod Sagar, Kothandam Sivakumar
Zitieren
APA
Mumtaz, Z. A., Sagar, P., Sivakumar, K., Mohakud, A. R., Rajendran, M., & Pavithran, S. (2023). Risk factors associated with device embolisation or malposition during transcatheter closure of patent ductus arteriosus. Cardiology in the Young, 33(10), 2041-2048. https://doi.org/10.1017/s1047951122003973
Vancouver
Mumtaz ZA, Sagar P, Sivakumar K, Mohakud AR, Rajendran M, Pavithran S. Risk factors associated with device embolisation or malposition during transcatheter closure of patent ductus arteriosus. Cardiology in the Young. 2023;33(10):2041-2048. doi:10.1017/s1047951122003973
BibTeX
@article{Mumtaz2023,
author = {Mumtaz, Zeeshan A and Sagar, Pramod and Sivakumar, Kothandam and Mohakud, Asish R and Rajendran, Monica and Pavithran, Sreeja},
title = {Risk factors associated with device embolisation or malposition during transcatheter closure of patent ductus arteriosus},
journal = {Cardiology in the Young},
year = {2023},
volume = {33},
number = {10},
pages = {2041--2048},
doi = {10.1017/s1047951122003973}
}
EUVIMED