EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Sustained radiation reduction following initial quality improvement intervention in a paediatric cardiac catheterisation laboratory

Anthony McKeiver iD, Jennifer Marshall, Peter Churchill, Douglas Bittel, James E O'Brien, Stephen Kaine, Michael Bingler

Cardiology in the Young · 2023-02 · 33 (2) · 221-226

DOI: 10.1017/s1047951122000543 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Cardiology in the Young, online erschienen am 18.03.2022. Verfasst von Anthony McKeiver und 6 weiteren Autor:innen. Beteiligte Einrichtung: College of Medicine, Kansas City University, Kansas City, MO, USA. Hauptthemen laut MEDLINE: Quality Improvement, Radiation Exposure. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

As part of a quality improvement project beginning in October 2011, our centre introduced changes to reduce radiation exposure during paediatric cardiac catheterisations. This led to significant initial decreases in radiation to patients. Starting in April 2016, we sought to determine whether these initial reductions were sustained.

Methods

After a 30-day trial period, we implemented (1) weight-based reductions in preset frame rates for fluoroscopy and angiography, (2) increased use of collimators and safety shields, (3) utilisation of stored fluoroscopy and virtual magnification, and (4) hiring of a devoted radiation technician. We collected patient weight (kg), total fluoroscopy time (min), and procedure radiation dosage (cGy-cm2) for cardiac catheterisations between October, 2011 and September, 2019.

Results

A total of 1889 procedures were evaluated (196 pre-intervention, 303 in the post-intervention time period, and 1400 in the long-term group). Fluoroscopy times (18.3 ± 13.6 pre; 19.8 ± 14.1 post; 17.11 ± 15.06 long-term, p = 0.782) were not significantly different between the three groups. Patient mean radiation dose per kilogram decreased significantly after the initial quality improvement intervention (39.7% reduction, p = 0.039) and was sustained over the long term (p = 0.043). Provider radiation exposure was also significantly decreased from the onset of this project through the long-term period (overall decrease of 73%, p < 0.01) despite several changes in the interventional cardiologists who made up the team over this time period.

Conclusion

Introduction of technical and clinical practice changes can result in a significant reduction in radiation exposure for patients and providers in a paediatric cardiac catheterisation laboratory. These reductions can be maintained over the long term.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Quality ImprovementRadiation ExposureChildHumansRadiation DosageAngiographyCardiac CatheterizationFluoroscopy

Stichwörter der Autor:innen

Catheterisationfluoroscopypaediatricquality improvementradiation exposure

Autor:innen und Institutionen

  • College of Medicine, Kansas City University, Kansas City, MO, USA.
  • Ward Family Heart Center, Children's Mercy Hospital, Kansas City, MO, USA.
  • Division of Cardiology, Department of Cardiovascular Services, Nemours Children's Hospital, Orlando, FL, USA.

ORCID-Profile: Anthony McKeiver

Zitieren

APA

McKeiver, A., Marshall, J., Churchill, P., Bittel, D., O'Brien, J. E., Kaine, S., & Bingler, M. (2023). Sustained radiation reduction following initial quality improvement intervention in a paediatric cardiac catheterisation laboratory. Cardiology in the Young, 33(2), 221-226. https://doi.org/10.1017/s1047951122000543

Vancouver

McKeiver A, Marshall J, Churchill P, Bittel D, O'Brien JE, Kaine S, et al. Sustained radiation reduction following initial quality improvement intervention in a paediatric cardiac catheterisation laboratory. Cardiology in the Young. 2023;33(2):221-226. doi:10.1017/s1047951122000543

BibTeX
@article{McKeiver2023,
  author = {McKeiver, Anthony and Marshall, Jennifer and Churchill, Peter and Bittel, Douglas and O'Brien, James E and Kaine, Stephen and Bingler, Michael},
  title = {Sustained radiation reduction following initial quality improvement intervention in a paediatric cardiac catheterisation laboratory},
  journal = {Cardiology in the Young},
  year = {2023},
  volume = {33},
  number = {2},
  pages = {221--226},
  doi = {10.1017/s1047951122000543}
}