Design and technical validation of a low-cost, titanium sheet metal TORP (SM-T) for manufacturing and use in LMICs
The Journal of Laryngology & Otology · 2026-05 · 140 (5) · 563-570
DOI: 10.1017/s0022215126104666 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift The Journal of Laryngology & Otology, online erschienen am 30.03.2026. Verfasst von Arjan J Knulst und 8 weiteren Autor:innen. Beteiligte Einrichtung: INF Green Pastures Hospital and Rehabilitation Center, International Nepal Fellowship, Nepal. Hauptthemen laut MEDLINE: Ossicular Prosthesis, Titanium, Prosthesis Design, Ossicular Replacement. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-SA 4.0).
Abstract (Originalsprache)
Objectives
Total ossicular replacement prostheses are commonly used to restore hearing in the conductive hearing impaired. However, total ossicular replacement prostheses are expensive and scarcely available in low- and middle-income countries. Therefore, the aim was to develop and validate an open-source design of a total ossicular replacement prosthesis that can be produced locally for low- and middle-income countries hospitals.
Methods
Based on input from literature and clinicians a design was made and verified. Accuracy, re-bending safety, cadaver ear fitting, sound transmission and production learning curves were tested.
Results
A titanium sheet metal total ossicular replacement prosthesis, costing 9 euro, showed accurate dimensions, surgeons showed short learning curves, and sound transmission was similar to commercial total ossicular replacement prostheses. Re-bending tests demonstrated bending corrections could be safely done. Good cadaver ear fitting could be demonstrated, without conflicts between the titanium sheet metal total ossicular replacement prosthesis and anatomical features.
Conclusion
The titanium sheet metal total ossicular replacement prosthesis design has proved to be feasible for implementation in low- and middle-income countries.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- INF Green Pastures Hospital and Rehabilitation Center, International Nepal Fellowship, Nepal.
- Department of Biomechanical Engineering, Delft University of Technologyhttps://ror.org/02e2c7k09, The Netherlands.
ORCID-Profile: Arjan J Knulst
Zitieren
APA
Knulst, A. J., van Asperen, V., Gautam, N. R., Geerts, R., Ten Have, G., Hou, S., Neuman, A., Smith, M., & Dankelman, J. (2026). Design and technical validation of a low-cost, titanium sheet metal TORP (SM-T) for manufacturing and use in LMICs. The Journal of Laryngology & Otology, 140(5), 563-570. https://doi.org/10.1017/s0022215126104666
Vancouver
Knulst AJ, van Asperen V, Gautam NR, Geerts R, Ten Have G, Hou S, et al. Design and technical validation of a low-cost, titanium sheet metal TORP (SM-T) for manufacturing and use in LMICs. The Journal of Laryngology & Otology. 2026;140(5):563-570. doi:10.1017/s0022215126104666
BibTeX
@article{Knulst2026,
author = {Knulst, Arjan J and van Asperen, Vivian and Gautam, Nabin Raj and Geerts, Robert and Ten Have, Geert and Hou, Sophia and Neuman, Alexandra and Smith, Michael and Dankelman, Jenny},
title = {Design and technical validation of a low-cost, titanium sheet metal TORP (SM-T) for manufacturing and use in LMICs},
journal = {The Journal of Laryngology & Otology},
year = {2026},
volume = {140},
number = {5},
pages = {563--570},
doi = {10.1017/s0022215126104666}
}
EUVIMED